Created: Jun 19, 2025
Due: Jan 31, 2026
Completed: Sep 30, 2025 (123 days early)
📅
71.0 daysTotal Age
⏱️
61.0d in Ready for SchedulingLongest Stage
🔄
9 transitionsStatus Changes
⚠️
61.2d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
4 linesWork Details
🏠
1 visitOn-Site Trips
📋
CoordinatorScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 258 days old
🚨
Primary Delay61.0 days stuck in "Ready for Scheduling"
⏱️
Total Delay61.2 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Release
11.0d
Ready for Scheduling
61.0d / 1.0d SLA
Scheduled
0.7d / 2.0d SLA
Closed
1.2d / 1.0d SLA
Ready to Bill
2.1d / 1.0d SLA
Invoice Pending
2.6d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
09/22/2025 RArnold- Called 347-346-1212 for Josiah James to schedule Alberto next week. Tenativly scheduling for 09/29. Will keep him updated on arrival date. WOs on demand.
07/07/2025 MHamersley - rec'd email reply from Ujasi Dalal, confirming all devices; POC: Josiah James, 347-346-1212, JosiahJames.Pangilinan@intelashealth.com ; FSE Credentialing: Check in at entrance to get pass; ... 07/03/2025 MHamersley - called 646-937-1865, call went to VM/LM; sent email: Ujasi.Dalal@Crothall.com ; RE: 19 Venue Go Ultrasound Devices
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R3.4 or R3.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
N/A
⚠️Problem Description
FMI 78101B: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.4 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📅
WO-01281645 - FCO - Venue Go Ultrasound - BROOKLYN - CONEY ISLAND HOSPITAL
Sep 30, 2025 8:00am – 5:00pm
Work Order: WO-01281645
Account: GE Healthcare FMI
Customer Case Number: 1-554879859869, 1-554646614122
Location: CONEY ISLAND HOSPITAL
Address: 2601 OCEAN PKWY, BROOKLYN, NY 11235
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGC006557
Contact Name: Josiah James
Contact Phone:
JPG
VGC006557 - PRODUCT LABEL
Sep 30, 2025
JPG
VGC006557 - DIAGNOSTIC TEST
Sep 30, 2025
JPG
VGC006557 - BATTERY REPLACEMENT
Sep 30, 2025
Timeline
📋
Work Order Created Jun 19, 2:26 PM
Work Order:WO-01281645
Type:FCO
Priority:PC
✅
Status: Ready for Release by Allison BelyeaJun 19, 3:16 PM
From:Entered
To:Ready for Release
Duration in Previous:50m
🚨11.0 business days
📅