Created: Jun 19, 2025
Due: Jan 31, 2026
Completed: Sep 7, 2025 (146 days early)
📅
54.0 daysTotal Age
⏱️
31.0d in ScheduledLongest Stage
🔄
15 transitionsStatus Changes
⚠️
45.9d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
8 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 258 days old
🚨
Primary Delay31.0 days stuck in "Scheduled"
⏱️
Total Delay45.9 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Release
24.0d
Scheduled
31.0d / 2.0d SLA
Tech On Site
1.0d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Incomplete
0.0d
Closed
3.0d / 1.0d SLA
Ready to Bill
18.3d / 1.0d SLA
Billing Review
0.0d / 2.0d SLA
Invoice Pending
2.6d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
07/24/2025 DMalloy - Called 216.312.6486 for Adrianna Payne. She agreed to have FSEs Paul/Jackson onsite beginning Monday 8/11/2025. Emailed & Received confirmation: paynea3@ccf.org ...
07/02/2025 MHamersley - called 216-312-6486, spoke to Adrianna, she is POC, sending her excel spreadsheet with all devices, she will go through and let us know which ones are hers and provide contacts for the others; sent email: PAYNEA3@ccf.org ; RE: 124 Venue Go Ultrasound Devices
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R4.4.3 and communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing new supplied batteries / Batteries installed by customer and dispose of old batteries according to local regulation”.
Verification Test:
“Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional”.
🔍Technician Findings
N/A
⚠️Problem Description
FMI 78101E: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.4.3 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Work Order: WO-01281910
Account: GE Healthcare FMI
Customer Case Number: 1-555020651136, 1-554790853286
Location: CLEVELAND CLINIC
Address: 2035 E 86TH ST, CLEVELAND, OH 44106
Manufacturer: GE Healthcare
Model #: Venue Fit Ultrasound
Serial #: VFC000513
Contact Name: Adrianna Payne (CE Team