📍 NEWARK WAYNE COMMUNITY HOSPITAL — 1200 DRIVING PARK AVE, NEWARK NY, 14513
Jun 19, 2025 → Feb 9, 2026
🔴 Behind
PC
SLA Clock
Per client agreement · Target: Feb 28, 2026
✓ SLA Met (12 biz days early)
Active: 2 biz days
Paused: 157 biz days
🕐 Clock stopped: Completed (Feb 10, 2026)
▶ Clock Running
Status: Created
Jun 19 → Jun 19, 2025
0 business days (counted)
Running total: 0 of ? biz days used
⏸157d
⏸ Clock Paused
Status: Ready for Release
Jun 19 → Feb 6, 2026
157 business days (not counted)
▶ Clock Running
Status: Scheduled
Feb 6 → Feb 9, 2026
1 business day (counted)
Running total: 1 of ? biz days used
▶ Clock Running
Status: Tech On Site
Feb 9 → Feb 10, 2026
1 business day (counted)
Running total: 2 of ? biz days used
Created: Jun 19, 2025Completed: Feb 10, 2026
Show SLA Clock Detail (4 periods)
Period
Status
Clock
Biz Days
Running Total
Jun 19 → Jun 19
Created
▶ Running
0
0
Jun 19 → Feb 6
Ready for Release
⏸ Paused
157
—
Feb 6 → Feb 9
Scheduled
▶ Running
1
1
Feb 9 → Feb 10
Tech On Site
▶ Running
1
2
Holding157.0d (excluded)
Coordinators1.6d
Field Work0.5d
✓ SLA Met
Holding(excluded from total)
157.0 days
Coordinators
1.6 days
Field Work
0.5 days
Created: Jun 19, 2025
Due: Feb 28, 2026 (5 days overdue)
📅
158.0 daysTotal Age
⏱️
157.0d in Ready for ReleaseLongest Stage
🔄
5 transitionsStatus Changes
⚠️
3.0d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
9 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
On Track
Ready to Bill • 258 days old
🚨
Primary Delay157.0 days stuck in "Ready for Release"
⏱️
Total Delay3.0 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Release
157.0d
Scheduled
1.6d / 2.0d SLA
Tech On Site
0.5d / 1.0d SLA
Closed
4.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/06/2026 DMalloy - Called 585.537.2756 phone number located for BioMed Randy Graves. Call not connecting. Emailed customer the recall info and advised Eric would be in the area on next week.
randy.graves@trimedx.com
julieta.mejia-trejos@rochesterregional.org
wallace.king@rochesterregional.org
01/19/2026 YEspino Called 315-332-2022 to verify the ultrasound device in question. Asked to be transferred over to the biomed dept, but then was advised that the line is busy and was advised to call back.
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: Completed FMI 78101 by installing SW Version R3.5 and communicated with customer to scrap previous software versions.
Verification Test: Performed Probe and Connectors Checks for all probes. Confirmed all probes pass.Completed all Peripheral Checks-system passed.Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
78101
⚠️Problem Description
FMI 78101C: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Fit Ultrasound
📞Call Description
FMI 78101
📧
Email: Urgent Schedule Attempt- Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - NEWARK WAYNE COMMUNITY HOSPITAL - NEWARK, NY - WO-01282144
Completed Feb 6, 2026
To: randy.graves@trimedx.com; julieta.mejia-trejos@rochesterregional.org; wallace.king@rochesterregional.org
CC: central.scheduling@emsar.com; fmessak@emsar.com; eric.mancerella@emsar.com; sean.currie@emsar.com
BCC: deon.malloy@emsar.com
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Urgent Schedule Attempt- Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - NEWARK WAYNE COMMUNITY HOSPITAL - NEWARK, N
📞
Live Call_Asked To Call Back
Completed Jan 19, 2026
01/19/2026 YEspino Called 315-332-2022 to verify the ultrasound device in question. Asked to be transferred over to the biomed dept, but then was advised that the line is busy and was advised to call back.
📅
WO-01282144 - FCO - Venue Fit Ultrasound - NEWARK - NEWARK WAYNE COMMUNITY HOSPITAL
Feb 9, 2026 8:00am – 5:00pm
Work Order: WO-01282144
Account: GE Healthcare FMI
Customer Case Number: 1-554880085976, 1-554907756161
Location: NEWARK WAYNE COMMUNITY HOSPITAL
Address: 1200 DRIVING PARK AVE, NEWARK, NY 14513
Manufacturer: GE Healthcare
Model #: Venue Fit Ultrasound
Serial #: VFC000839
Contact Name: Randy
PDF
VFC000839 COMPLETION REPORT
Feb 9, 2026
JPG
VFC000839 WORKLIST NOT CONNECTED TO WIFI
Feb 9, 2026
JPG
VFC000839 BATTERY
Feb 9, 2026
JPG
VFC000839 UNIT NOT CONNECTED TO WIFI
Feb 9, 2026
JPG
VFC000839 SERIAL #
Feb 9, 2026
JPG
VFC000839 SOFTWARE
Feb 9, 2026
Timeline
📋
Work Order Created Jun 19, 2:27 PM
Work Order:WO-01282144
Type:FCO
Priority:PC
✅
Status: Ready for Release by Allison BelyeaJun 19, 3:14 PM