Created: Oct 3, 2025
Due: Jan 31, 2026
Completed: Dec 4, 2025 (58 days early)
📅
40.3 daysTotal Age
⏱️
37.3d in EnteredLongest Stage
🔄
12 transitionsStatus Changes
⚠️
55.3d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
11 linesWork Details
🏠
1 visitOn-Site Trips
📋
CoordinatorScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 152 days old
🚨
Primary Delay37.3 days stuck in "Entered"
⏱️
Total Delay55.3 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
37.3d / 0.5d SLA
Scheduled
4.7d / 2.0d SLA
Closed
1.0d / 1.0d SLA
Ready to Bill
21.5d / 1.0d SLA
Billing Review
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
12/01/2025 Brod - Fady emailed 'JosiahJames.Pangilinan@intelashealth.com' to have FSE onsite for 12/04. Josiah confirmed
10/03/2025 SRobinson- Created follow up WO as requested. (Old WO-01281236) -Follow up notes - BR 10/01/2025 - Please create a follow-up EWO for this serial # due to issuses found prior to FMI
-Previous notes -
09/22/2025 RArnold- Called 347-346-1212 for Josiah James to schedule Alberto next week. Tenativly scheduling for 09/29. Will keep him updated on arrival date. WOs on demand. 07/07/2025 MHamersley - rec'd email reply from Ujasi Dalal, confirming all devices; POC: Josiah James, 347-346-1212, JosiahJames.Pangilinan@intelashealth.com ; FSE Credentialing: Check in at entrance to get pass; ... 07/03/2025 MHamersley - called 646-937-1865, call went to VM/LM; sent email: Ujasi.Dalal@Crothall.com ; RE: 19 Venue Go Ultrasound Devices
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R3.4 and communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing new supplied batteries / Batteries installed by customer and dispose of old batteries according to local regulation”.
Verification Test:
🔍Technician Findings
N/A
⚠️Problem Description
FMI 78101B: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.4 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📅
WO-01307751 - FCO - Venue Go Ultrasound - BROOKLYN - CONEY ISLAND HOSPITAL
Dec 3, 2025 7:00pm – 7:00pm
Work Order: WO-01307751
Account: GE Healthcare FMI
Customer Case Number: 1-554879871248, 1-554907740802
Location: CONEY ISLAND HOSPITAL
Address: 2601 OCEAN PKWY, BROOKLYN, NY 11235
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGC006568
Contact Name: Josiah James