Work Order 360 View

Detailed work order timeline, KPIs, and analysis

← WO Overview

WO-01315590 ↗ ServiceMax

GE HEALTHCARE • Repair • P5

📍 St Francis Hospital — 2122 Manchester Expressway, Columbus GA, 31904

Nov 17, 2025 → Nov 17, 2025

✅ Completed
P5
SLA Clock
5 business days from creation · Target: Nov 24, 2025
✓ SLA Met (5 biz days early)
Active: 0 biz days
Budget: 5 biz days
🕐 Clock stopped: Completed (Nov 17, 2025)
SLA Target
▶ Clock Running
Status: Created
Nov 17 → Nov 17, 2025
0 business days (counted)
Running total: 0 of 5 biz days used
0d
▶ Clock Running
Status: Ready for Scheduling
Nov 17 → Nov 17, 2025
0 business days (counted)
Running total: 0 of 5 biz days used
0d
▶ Clock Running
Status: Assigned
Nov 17 → Nov 17, 2025
0 business days (counted)
Running total: 0 of 5 biz days used
Created: Nov 17, 2025 Completed: Nov 17, 2025
Business Days Used 0 / 5
0 5d budget
Show SLA Clock Detail (3 periods)
Period Status Clock Biz Days Running Total
Nov 17 → Nov 17 Created ▶ Running 0 0 / 5
Nov 17 → Nov 17 Ready for Scheduling ▶ Running 0 0 / 5
Nov 17 → Nov 17 Assigned ▶ Running 0 0 / 5
Dispatch 0.0d
Coordinators 0.3d
Billing 0.1d
✓ SLA Met
Dispatch
0.0 days
Coordinators
0.3 days
Billing
0.1 days
Created: Nov 17, 2025 Due: Nov 24, 2025 Completed: Nov 17, 2025 (7 days early)
📅
0.3 days Total Age
⏱️
1.5d in Closed Longest Stage
🔄
5 transitions Status Changes
⚠️
0.5d over Total Delay
📦
0 orders, 0 quotes Parts & Quotes
📝
2 lines Work Details
🏠
1 visit On-Site Trips
📋
Coordinator Scheduling

AI Analysis

Powered by OpenAI
Generating AI analysis...
Completed with minor delays
Invoiced • 107 days old
🚨
Primary Delay 1.5 days stuck in "Closed"
⏱️
Total Delay 0.5 business days over SLA thresholds

SLA vs Actual Time by Status

Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
0.1d / 1.0d SLA
Assigned
0.2d / 0.5d SLA
Closed
1.5d / 1.0d SLA
Ready to Bill
0.1d / 1.0d SLA

💬 Activities, Notes & Communications

🔧 Work Performed
performed calibration on bed tested all functions and returned bed to service
🔍 Technician Findings
found bed alarm was triggered thru IBED awareness due to bed reading fowler angles incorrectly and required bed to be calibrated to correct fault
⚠️ Problem Description
Requested by Kenneth Lancara: repair request bed alarm goes off when disarmed
📄 Description
Stryker FL27 InTouch (series) Critical Care Bed (2130, 2131, 2040, 2141, 2151, 2152, 2153, 2154, 2155, 2156)

Timeline

📋
2 Work Order Created Nov 17, 1:10 PM
Work Order Created → Status: Ready for Scheduling
📅
Status: Ready for Scheduling by Kimberly Hardison
From: Entered
To: Ready for Scheduling
Duration in Previous: 0m
👷
2 Technician Assigned Nov 17, 2:00 PM
Technician Assigned → Status: Assigned
👷
Status: Assigned by Abby Jinerson
From: Ready for Scheduling
To: Assigned
Duration in Previous: 49m
⏱️
2 2x Labor Added Nov 17, 3:18 PM
2x Labor Added → Status: Completed
✔️
Status: Completed by Kenneth Lancara
From: Assigned, Completed
To: Completed, Closed
Duration in Previous: 1h 18m, 0m
📄
Status: Ready to Bill by Hannah Shaw Nov 18, 1:08 PM
From: Closed
To: Ready to Bill
Duration in Previous: 21h 49m
📨
Status: Invoice Pending by Hannah Shaw Nov 18, 2:06 PM
From: Ready to Bill, Invoice Pending
To: Invoice Pending, Invoiced
Duration in Previous: 58m, 0m
2.0 business days
🧾
Invoice Created Nov 19, 9:24 PM
Invoice Number: SM-246877

Bottlenecks

Closed
Duration: 1.5 business days (threshold: 1.0 business days)
Exceeded by 3.7 business hours

🔗 Related Artifacts

No related artifacts

ℹ️ Work Order Details

Status
Invoiced
Priority
P5
Type
Repair
Branch
N/A
Technician
Kenneth Lancara
Scheduled
N/A
Created
Nov 17, 1:10 PM
Invoice #
SM-246877

Work Details (2)

Name Type Status Qty Amount Date
WL-03010102 Labor Open 0.5 $0.00 Nov 17, 3:18 PM
WL-03010103 Labor Open 2 $0.00 Nov 17, 3:18 PM

Details