5 business days from creation · Target: Dec 1, 2025
✓ SLA Met (3 biz days early)
Active: 2 biz days
Budget: 5 biz days
🕐 Clock stopped: Completed (Nov 24, 2025)
SLA Target
▶ Clock Running
Status: Created
Nov 20 → Nov 20, 2025
0 business days (counted)
Running total: 0 of 5 biz days used
▶ Clock Running
Status: Ready for Scheduling
Nov 20 → Nov 20, 2025
0 business days (counted)
Running total: 0 of 5 biz days used
2d
▶ Clock Running
Status: Assigned
Nov 20 → Nov 24, 2025
2 business days (counted)
Running total: 2 of 5 biz days used
Created: Nov 20, 2025Completed: Nov 24, 2025
Business Days Used2 / 5
05d budget
Show SLA Clock Detail (3 periods)
Period
Status
Clock
Biz Days
Running Total
Nov 20 → Nov 20
Created
▶ Running
0
0 / 5
Nov 20 → Nov 20
Ready for Scheduling
▶ Running
0
0 / 5
Nov 20 → Nov 24
Assigned
▶ Running
2
2 / 5
Dispatch0.0d
Coordinators2.4d
Billing0.2d
✓ SLA Met
Dispatch
0.0 days
Coordinators
2.4 days
Over SLA
Billing
0.2 days
Created: Nov 20, 2025
Due: Dec 1, 2025
Completed: Nov 24, 2025 (7 days early)
📅
2.4 daysTotal Age
⏱️
2.4d in AssignedLongest Stage
🔄
5 transitionsStatus Changes
⚠️
2.2d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
1 linesWork Details
🏠
1 visitOn-Site Trips
📋
CoordinatorScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with minor delays
Invoiced • 104 days old
🚨
Primary Delay2.4 days stuck in "Assigned"
⏱️
Total Delay2.2 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
0.0d / 1.0d SLA
Assigned
2.4d / 0.5d SLA
Closed
1.3d / 1.0d SLA
Ready to Bill
0.2d / 1.0d SLA
💬 Activities, Notes & Communications
🔧Work Performed
11.25.2025
The bed alarm issue was corrected by recalibrating the scale system in accordance with Stryker manufacturer specifications. All calibration steps were completed successfully, and the scale and bed-exit functions were reinitialized. A full operational check was performed, and the alarm system is now responding correctly.
🔍Technician Findings
Bed alarm was not functioning due to the scales being out of calibration. Recalibration was performed per Stryker guidelines, restoring proper scale accuracy and bed-exit alarm functionality. No additional faults were observed after calibration.
⚠️Problem Description
11.20.2025 Note on unit states "Bed alarm not working".
Diag needed to determine root-cause