📍 DELL SETON MEDICAL CENTER AT THE UNIVERSITY OF TEXAS — 1500 RED RIVER ST, AUSTIN TX, 78701
Dec 11, 2025 → Dec 19, 2025
✅ Completed
PC
SLA Clock
Per client agreement · Target: Jan 31, 2026
✓ SLA Met (27 biz days early)
Active: 4 biz days
Paused: 2 biz days
🕐 Clock stopped: Completed (Dec 19, 2025)
1d
▶ Clock Running
Status: Created
Dec 11 → Dec 12, 2025
1 business day (counted)
Running total: 1 of ? biz days used
▶ Clock Running
Status: Assigned
Dec 12 → Dec 12, 2025
0 business days (counted)
Running total: 1 of ? biz days used
▶ Clock Running
Status: Scheduled
Dec 12 → Dec 13, 2025
0 business days (counted)
Running total: 1 of ? biz days used
▶ Clock Running
Status: Tech On Site
Dec 13 → Dec 13, 2025
0 business days (counted)
Running total: 1 of ? biz days used
3d
▶ Clock Running
Status: Reschedule
Dec 13 → Dec 17, 2025
3 business days (counted)
Running total: 4 of ? biz days used
⏸2d
⏸ Clock Paused
Status: Ready for Release
Dec 17 → Dec 19, 2025
2 business days (not counted)
▶ Clock Running
Status: Tech On Site
Dec 19 → Dec 19, 2025
0 business days (counted)
Running total: 4 of ? biz days used
Created: Dec 11, 2025Completed: Dec 19, 2025
Show SLA Clock Detail (7 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 11 → Dec 12
Created
▶ Running
1
1
Dec 12 → Dec 12
Assigned
▶ Running
0
1
Dec 12 → Dec 13
Scheduled
▶ Running
0
1
Dec 13 → Dec 13
Tech On Site
▶ Running
0
1
Dec 13 → Dec 17
Reschedule
▶ Running
3
4
Dec 17 → Dec 19
Ready for Release
⏸ Paused
2
—
Dec 19 → Dec 19
Tech On Site
▶ Running
0
4
Dispatch2.0d
Holding2.0d (excluded)
Coordinators4.0d
Field Work1.0d
Billing2.5d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
2.0 days
Coordinators
4.0 days
Over SLA
Field Work
1.0 days
Billing
2.5 days
Over SLA
Created: Dec 11, 2025
Due: Jan 31, 2026
Completed: Dec 19, 2025 (43 days early)
📅
7.0 daysTotal Age
⏱️
6.5d in ClosedLongest Stage
🔄
9 transitionsStatus Changes
⚠️
9.5d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
13 linesWork Details
🏠
2 visitsOn-Site Trips
👤
Self-ScheduledScheduling
🔄
2 On-Site Visits
(1 reschedule)
2 on-site visits — follow-up trip required
👤 Self-scheduled by Marvin Rivas — scheduling delays attributed to technician, not coordinators
Visit 1Dec 12, 2025
Visit 2Dec 18, 2025
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with minor delays
Invoiced • 83 days old
🚨
Primary Delay6.5 days stuck in "Closed"
⏱️
Total Delay9.5 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Scheduled
1.0d / 2.0d SLA
Tech On Site
1.0d / 1.0d SLA
Reschedule
3.0d / 2.0d SLA
Ready for Release
2.0d
Completed
0.0d / 1.0d SLA
Closed
6.5d / 1.0d SLA
Ready to Bill
2.5d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
12/18/2025 FSE Marvin Rivas - SW update only. Batteries will be installed by customer upon delivery.
12/12/2025 FSE Marvin Rivas – incorrect software on hands, and no batteries on arrival.
12/12/2025 MNewman - Dispatching to Marvin per Fady for today 12/12/2025.
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R4.4.3 or R4.5 and communicated with
customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed all probes pass.
Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully
functional”.
🔍Technician Findings
FMI software update completed. Device returned to use.
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.4.3 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.