📍 ED FRASER MEMORIAL HOSPITAL — 159 N 3RD ST, MACCLENNY FL, 32063
Dec 12, 2025 → Feb 2, 2026
🔴 Behind
PC
SLA Clock
Per client agreement · Target: Jan 31, 2026
✗ SLA Missed
Active: 3 biz days
Paused: 31 biz days
🕐 Clock stopped: Completed (Feb 3, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸31d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Jan 30, 2026
31 business days (not counted)
▶ Clock Running
Status: Assigned
Jan 30 → Jan 30, 2026
0 business days (counted)
Running total: 1 of ? biz days used
▶ Clock Running
Status: Scheduled
Jan 30 → Feb 2, 2026
1 business day (counted)
Running total: 2 of ? biz days used
▶ Clock Running
Status: Tech On Site
Feb 2 → Feb 3, 2026
1 business day (counted)
Running total: 3 of ? biz days used
Created: Dec 12, 2025Completed: Feb 3, 2026
Show SLA Clock Detail (5 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 12 → Dec 15
Created
▶ Running
1
1
Dec 15 → Jan 30
Ready for Release
⏸ Paused
31
—
Jan 30 → Jan 30
Assigned
▶ Running
0
1
Jan 30 → Feb 2
Scheduled
▶ Running
1
2
Feb 2 → Feb 3
Tech On Site
▶ Running
1
3
Dispatch2.0d
Holding31.9d (excluded)
Coordinators2.0d
Field Work1.0d
✗ SLA Missed
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
31.9 days
Coordinators
2.0 days
Field Work
1.0 days
Created: Dec 12, 2025
Due: Jan 31, 2026 (33 days overdue)
📅
34.2 daysTotal Age
⏱️
31.9d in Ready for ReleaseLongest Stage
🔄
5 transitionsStatus Changes
⚠️
6.5d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
7 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
At Risk - Monitor Closely
Ready to Bill • 82 days old
🚨
Primary Delay31.9 days stuck in "Ready for Release"
⏱️
Total Delay6.5 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
31.9d
Scheduled
2.0d / 2.0d SLA
Tech On Site
1.0d / 1.0d SLA
Closed
6.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
01/29/2026 ALuman - received call from Kevin Doyle and he confirmed device onsite. Dispatch to Jared Heisler to complete on 2/2. Email confirmation sent to kdoyle@bcmedsvcs.com
01/29/2026 MHamersley - called 904-259-3151, ext 2254; call went to VM/LM;
01/29/2026 ALuman - Fady emailed customer via existing thread with urgent notice to complete recall. Waiting for response
01/23/2026 YEspino Called 904-259-3151 to speak to Geoffry @ Ext 2254 to verify the device in question. Was transferred over to Respt and was advised to call back because hes not in.
01/22/2026 ALuman - sent follow up email to Geoffry Moran, looping in Fady asking to confirm and schedule
01/13/2026 ALuman - called Geoffry Moran 904-259-3151 ext 2254. No answer, left a partial VM before system cut it off and hung up automatically. Sending email to GMoran@bcmedsvcs.com
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: Completed FMI 78101 by installing SW Version R4.5 and communicated with customer to scrap previous software versions.
Verification Test: Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
"Template Name:
Template Description:
Problem Found:
Action Taken: Completed FMI 78101 by installing SW Version R4.5 and communicated with customer to scrap previous software versions.
Verification Test: Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
⚠️Problem Description
FMI 78101E: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Fit Ultrasound
📞Call Description
FMI 78101
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - ED FRASER MEMORIAL HOSPITAL - MACCLENNY, FL - WO-01325467
Completed Jan 29, 2026
To: kdoyle@bcmedsvcs.com; jared.heisler@emsar.com; fmessak@emsar.com
CC: central.scheduling@emsar.com; douglas.pearce@emsar.com
BCC: ashley.luman@emsar.com
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - ED FRASER MEMORIAL HOSPITAL - MACCLENNY, FL - WO-01325467
Body:
Hi Kevin,
Thank you for taking my call today. As discussed, I am contacting you on b
📞
Live Call_Left Message
Completed Jan 29, 2026
01/29/2026 MHamersley - called 904-259-3151, ext 2254; call went to VM/LM;
📞
Live Call_Asked To Call Back
Completed Jan 23, 2026
01/23/2026 YEspino Called 904-259-3151 to speak to Geoffry @ Ext 2254 to verify the device in question. Was transferred over to Respt and was advised to call back because hes not in.
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - ED FRASER MEMORIAL HOSPITAL - MACCLENNY, FL - WO-01325467
Completed Jan 13, 2026
To: GMoran@bcmedsvcs.com
CC: central.scheduling@emsar.com
BCC: ashley.luman@emsar.com
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - ED FRASER MEMORIAL HOSPITAL - MACCLENNY, FL - WO-01325467
Body:
Hi Geoffry
I am contacting you on behalf of GE Healthcare about a Field Safety Notice issued for certain Venue Go and Venue Fit Ultrasound devices. The in
📅
WO-01325467 - FCO - Venue Fit Ultrasound - MACCLENNY - ED FRASER MEMORIAL HOSPITAL
Feb 2, 2026 8:00am – 5:00pm
Work Order: WO-01325467
Account: GE Healthcare FMI
Customer Case Number: 1-554880664264, 1-554908147993
Location: ED FRASER MEMORIAL HOSPITAL
Address: 159 N 3RD ST, MACCLENNY, FL 32063
Manufacturer: GE Healthcare
Model #: Venue Fit Ultrasound
Serial #: RVFD000930
Contact Name: Kevin Doyle
Contact Ph
JPEG
VFD000930-Battery
Feb 2, 2026
JPEG
VFD000930-Product Label Pic
Feb 2, 2026
JPEG
VFD000930-Software Update
Feb 2, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325467
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:10 PM