Created: Dec 12, 2025
Due: Jan 31, 2026 (33 days overdue)
📅
38.3 daysTotal Age
⏱️
30.9d in Ready for ReleaseLongest Stage
🔄
7 transitionsStatus Changes
⚠️
5.6d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
5 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
At Risk - Monitor Closely
Ready to Bill • 82 days old
🚨
Primary Delay30.9 days stuck in "Ready for Release"
⏱️
Total Delay5.6 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
30.9d
Scheduled
2.5d / 2.0d SLA
Tech On Site
1.4d / 1.0d SLA
Tech Off Site
4.2d / 2.0d SLA
Completed
0.0d / 1.0d SLA
Closed
2.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
01/29/2026 ALuman - dispatch to Konstantin Barsky to locate and complete FMI 1/30
01/26/2026 ALuman - followed up with another email to Johnathan Street, looping in Fady
01/13/2026 Luman - called Jonathan Street 863-314-5718. No answer, left VM. Sent email to Jonathan.Street@gehealthcare.com
01/12/2026 MHamersley - called 863-385-6101, xfer to Bio Med, call went to Johnathan's VM/LM
12/26/2025 MHamersley - called 863-385-6101, xfer to Bio Med, call went to Johnathan Straight's VM/LM
WO-01325504-Venue Go Ultrasound VGB001002
WO-01325505-Venue Go Ultrasound VGD002796
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R4.4.3 or R4.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
🔍Technician Findings
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R4.4.3 or R4.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Work Order: WO-01325505
Account: GE Healthcare FMI
Customer Case Number: 1-554880365932, 1-554908083982
Location: HCA FLORIDA HIGHLANDS HOSPITAL
Address: 3600 S HIGHLANDS AVE, SEBRING, FL 33870
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGD002796
Contact Name: Johnathan Stree
JPG
VGD002796_Product Label Pic
Feb 8, 2026
JPG
VGD002796_Product Pic.
Feb 8, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325505
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:09 PM