Created: Dec 12, 2025
Due: Feb 28, 2026 (5 days overdue)
📅
50.0 daysTotal Age
⏱️
41.9d in Ready for ReleaseLongest Stage
🔄
10 transitionsStatus Changes
⚠️
7.1d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
2 linesWork Details
🏠
1 visitOn-Site Trips
📋
CoordinatorScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
At Risk - Monitor Closely
Ready to Bill • 82 days old
🚨
Primary Delay41.9 days stuck in "Ready for Release"
⏱️
Total Delay7.1 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
41.9d
Scheduled
8.7d / 2.0d SLA
Tech On Site
0.0d / 1.0d SLA
Tech Off Site
0.0d / 2.0d SLA
Closed
2.0d / 1.0d SLA
Incomplete
0.0d
💬 Activities, Notes & Communications
🔒Internal CSA Notes
2/27/2026 BBallard - Device was located and completed. No follow up is needed. Pictures attached.
2/13/2026 Brod - emailed alan to have FSE onsite for 02/23. he confirmed for the 16th but the fse was no longer available.
02/12/2026 MHamersley - recd email reply from Sterlin, he provided: Alan Graves, alan.graves@hhsys.org , he is Bio Med for Ultrasounds.
02/12/2026 MHamersley - sent email: sterlin.woods@hhsys.org ;
*Email to jeffrey.stephens@hhsys.org Bounced*
WO-01325650-VFD000297
WO-01337755-VFC000952
12/26/2025 EMartin sent email to
Jeff Stephens 2562658893 jeffrey.stephens@hhsys.org and
Curtis Montgomery 800 626 4515 ext 113 cmontgomery@bmesco.com for update on device
🔧Work Performed
VFD000297 ANR
⚠️Problem Description
FMI 78101E: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Fit Ultrasound
📞Call Description
FMI 78101
📧
Email
Completed Feb 12, 2026
02/12/2026 MHamersley - sent email: sterlin.woods@hhsys.org ;
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - HUNTSVILLE HOSPITAL - HUNTSVILLE, AL - WO-01325650
Completed Feb 12, 2026
To: sterlin.woods@hhsys.org
CC: central.scheduling@emsar.com; brandi.rodriguez@emsar.com; mary.hamersley@emsar.com; fmessak@emsar.com
BCC:
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - HUNTSVILLE HOSPITAL - HUNTSVILLE, AL - WO-01325650
Body:
Good Afternoon Sterlin,
I am contacting you on behalf of GE Healthcare about a Field Safety Notice issued
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - HUNTSVILLE HOSPITAL - HUNTSVILLE, AL - WO-01325650
Completed Dec 26, 2025
To: jeffrey.stephens@hhsys.org; cmontgomery@bmesco.com
CC: erin.martin@emsar.com; central.scheduling@emsar.com
BCC:
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - HUNTSVILLE HOSPITAL - HUNTSVILLE, AL - WO-01325650
Body:
Hello,
I am contacting you on behalf of GE Healthcare about a Field Safety Notice issued for certain Venue Go and Venue Fit Ultr
Work Order: WO-01325650
Account: GE Healthcare FMI
Customer Case Number: 1-554880680183, 1-554908158087
Location: HUNTSVILLE HOSPITAL
Address: 101 SIVLEY RD SW, HUNTSVILLE, AL 35801
Manufacturer: GE Healthcare
Model #: Venue Fit Ultrasound
Serial #: VFD000297
Contact Name: Alan Graves
Contact Phone:
JFIF
VFD000297 SW
Feb 27, 2026
JFIF
VFD000297 SN
Feb 27, 2026
JFIF
VFD000297 Bat
Feb 27, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325650
Type:FCO
Priority:PC
⏳2.0 business days
✅