Created: Dec 12, 2025
Due: Jan 31, 2026
Completed: Jan 23, 2026 (8 days early)
📅
28.0 daysTotal Age
⏱️
24.9d in Ready for ReleaseLongest Stage
🔄
8 transitionsStatus Changes
⚠️
10.0d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
3 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 82 days old
🚨
Primary Delay24.9 days stuck in "Ready for Release"
⏱️
Total Delay10.0 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
24.9d
Scheduled
0.7d / 2.0d SLA
Tech On Site
2.2d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
6.4d / 1.0d SLA
Ready to Bill
2.9d / 1.0d SLA
Invoice Pending
1.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
01/07/2026 Cever called called 863-687-1100 x2804, phone just rings, no answer; unable to lvm will send email to wayne.buning@mylrh.org
01/07/2026 Cever called 8636871100 ext 5881 unable to reach anyone phone just rings unable to lvm will try another number
12/31/2025 MHamersley - called 863-687-1100 x2804, phone just rings, no answer; called main number, xfer to Bio Med, goes to VM not able to leave msg; sent f/u email: wayne.buning@mylrh.org ... 12/22/2025 Cever sent email to wayne.buning@mylrh.org trying to confirm inventory.
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R5.0.5 and
communicated with customer
to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed
all probes pass.
Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the
customer fully
functional”"
🔍Technician Findings
"Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R5.0.5 and
communicated with customer
to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed
all probes pass.
Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the
customer fully
functional”"
⚠️Problem Description
FMI 78101G: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R5.1 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📅
WO-01325675 - FCO - Venue Go Ultrasound - LAKELAND - LAKELAND REGIONAL MEDICAL
Jan 21, 2026 8:00am – 5:00pm
Work Order: WO-01325675
Account: GE Healthcare FMI
Customer Case Number: 1-554969316065
Location: LAKELAND REGIONAL MEDICAL
Address: 1324 LAKELAND HILLS BLVD TEST, LAKELAND, FL 33805
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGE000432
Contact Name:
Contact Phone:
Contact E
JPG
VGE000432_Product Label Pic
Jan 23, 2026
JPG
VGE000432_Software Update
Jan 23, 2026
JPG
VGE000432_Batteries
Jan 23, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325675
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:10 PM