Created: Dec 12, 2025
Due: Feb 28, 2026 (5 days overdue)
📅
51.0 daysTotal Age
⏱️
45.8d in Ready for ReleaseLongest Stage
🔄
6 transitionsStatus Changes
⚠️
4.1d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
3 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
On Track
Ready to Bill • 82 days old
🚨
Primary Delay45.8 days stuck in "Ready for Release"
⏱️
Total Delay4.1 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
45.8d
Scheduled
4.6d / 2.0d SLA
Tech On Site
0.5d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
1.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/27/2026 SRobinson- Created follow up WO as requested. (WO-01354236)
02/20/2026 Brod - called 9012875437 to speak to Barry, he stated to follow up in email with the list of units but confirmed FSE onsite for 02/25-02/26.
02/19/2026: ARoosa - Sent FU email to barry.jones@mlh.org, asking him to pls. advise on devices and locations
02/10/2026; ARoosa - Called; 901-287-5437, Oper Xfer to Biomed; Spoke to Barry who I spoke with earlier today he has verified he is contact for any methodist hospital
locations in Memphis as well as LE Bonheur children's. Sending Email with all locations
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R4.4.3 or R4.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
GEHC FMI 78101 WOM Process
Published: 06/12/2025 Page 7 of 14 v 1.0
Effective: 06/12/2025
Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
🔍Technician Findings
software updated
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📋
Outbound to +19012875437
Completed Feb 10, 2026
02/10/2026; ARoosa - Called; 901-287-5437, Oper Xfer to Biomed; Spoke to Barry who I spoke with earlier today he has verified he is contact for any methodist h
locations in Memphis as well as LE Bonheur children's. Sending Email with all locations
📅
WO-01325685 - FCO - Venue Go Ultrasound - MEMPHIS - LE BONHEUR CHILDRENS
Feb 25, 2026 7:00pm – 7:00pm
Work Order: WO-01325685
Account: GE Healthcare FMI
Customer Case Number: 1-554880300054, 1-554908014681
Location: LE BONHEUR CHILDRENS
Address: 848 ADAMS AVE, MEMPHIS, TN 38103
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGD000318
Contact Name: Barry Jones
Contact Phone: (901)
JPG
VGD000318_SN
Feb 26, 2026
JPG
VGD000318_Batteries
Feb 26, 2026
JPG
VGD000318_Software
Feb 26, 2026
Timeline
📋
Work Order Created Dec 12, 11:06 AM
Work Order:WO-01325685
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:14 PM