Work Order 360 View

Detailed work order timeline, KPIs, and analysis

← WO Overview

WO-01325958 ↗ ServiceMax

GE Healthcare FMI • FCO • PC

📍 PARKLAND HEALTH & HOSPITAL SYSTEM — 5223 TEX OAK AVE, DALLAS TX, 75235

Dec 12, 2025 → Feb 23, 2026

🔴 Behind
PC
SLA Clock
Per client agreement · Target: Feb 28, 2026
✓ SLA Met (4 biz days early)
Active: 8 biz days
Paused: 39 biz days
🕐 Clock stopped: Completed (Feb 23, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸39d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Feb 11, 2026
39 business days (not counted)
▶ Clock Running
Status: Assigned
Feb 11 → Feb 13, 2026
2 business days (counted)
Running total: 3 of ? biz days used
3d
▶ Clock Running
Status: Scheduled
Feb 13 → Feb 19, 2026
3 business days (counted)
Running total: 6 of ? biz days used
▶ Clock Running
Status: Completed
Feb 19 → Feb 20, 2026
1 business day (counted)
Running total: 7 of ? biz days used
▶ Clock Running
Status: Scheduled
Feb 20 → Feb 22, 2026
0 business days (counted)
Running total: 7 of ? biz days used
▶ Clock Running
Status: Completed
Feb 22 → Feb 22, 2026
0 business days (counted)
Running total: 7 of ? biz days used
▶ Clock Running
Status: Closed
Feb 22 → Feb 23, 2026
1 business day (counted)
Running total: 8 of ? biz days used
▶ Clock Running
Status: Incomplete
Feb 23 → Feb 23, 2026
0 business days (counted)
Running total: 8 of ? biz days used
Created: Dec 12, 2025 Completed: Feb 23, 2026
Show SLA Clock Detail (9 periods)
Period Status Clock Biz Days Running Total
Dec 12 → Dec 15 Created ▶ Running 1 1
Dec 15 → Feb 11 Ready for Release ⏸ Paused 39
Feb 11 → Feb 13 Assigned ▶ Running 2 3
Feb 13 → Feb 19 Scheduled ▶ Running 3 6
Feb 19 → Feb 20 Completed ▶ Running 1 7
Feb 20 → Feb 22 Scheduled ▶ Running 0 7
Feb 22 → Feb 22 Completed ▶ Running 0 7
Feb 22 → Feb 23 Closed ▶ Running 1 8
Feb 23 → Feb 23 Incomplete ▶ Running 0 8
Dispatch 2.0d
Holding 39.8d (excluded)
Coordinators 7.8d
Field Work 1.1d
Incomplete 0.0d (excluded)
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding (excluded from total)
39.8 days
Coordinators
7.8 days
Over SLA
Field Work
1.1 days
Over SLA
Incomplete (excluded from total)
0.0 days
Created: Dec 12, 2025 Due: Feb 28, 2026 (5 days overdue)
📅
48.0 days Total Age
⏱️
39.8d in Ready for Release Longest Stage
🔄
10 transitions Status Changes
⚠️
6.9d over Total Delay
📦
0 orders, 0 quotes Parts & Quotes
📝
5 lines Work Details
🏠
1 visit On-Site Trips
👤
Self-Scheduled Scheduling

AI Analysis

Powered by OpenAI
Generating AI analysis...
At Risk - Monitor Closely
Ready to Bill • 82 days old
🚨
Primary Delay 39.8 days stuck in "Ready for Release"
⏱️
Total Delay 6.9 business days over SLA thresholds

SLA vs Actual Time by Status

Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
39.8d
Assigned
2.8d / 0.5d SLA
Scheduled
5.0d / 2.0d SLA
Completed
1.1d / 1.0d SLA
Closed
2.9d / 1.0d SLA
Incomplete
0.0d

💬 Activities, Notes & Communications

🔒 Internal CSA Notes

02/23/2026 SRobinson- Created follow up WO as requested. (WO-01353210)

02/12/2026 mklose dispatch to Davon per Fady 02/10/2026 LHill-sent email to Dwayne Connor-trying to schedule tech visit for 14 devices. See logged email on WO-01325956 02/10/2026 LHill-Called 469-419-0375. Left message for Dwayne Connor . Trying to schedule tech visit for about 14 devices. See logged call on WO-01325956. 12/29/2025 MHamersley - recd email from Patrick providing contact information and cc in email: Dwayne Connor, Radiology Service Engineer, 469-419-0375; DWAYNE.CONNOR@phhs.org; ... 12/29/2025 MHamersley - called 214-590-8000 xfer to 5 different depts, spoke to Trisha Diaz, she said to contact: Patrick Eaton, 469-419-0274; email: patrick.eaton@phhs.org; called left VM and sent email.

🔧 Work Performed
Template Name: Template Description: Problem Found: Action Taken: “Completed FMI 78101 by installing SW Version R3.4 or R3.5 and communicated with customer to scrap previous software versions”. Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”. Template Name: Template Description: Problem Found: Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by customer and dispose of old batteries according to local regulation”. Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
🔍 Technician Findings
Template Name: Template Description: Problem Found: Action Taken: “Completed FMI 78101 by installing SW Version R3.4 or R3.5 and communicated with customer to scrap previous software versions”. Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”. Template Name: Template Description: Problem Found: Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by customer and dispose of old batteries according to local regulation”. Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
⚠️ Problem Description
FMI 78101B: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📞 Call Description
FMI 78101

Timeline

📋
Work Order Created Dec 12, 11:06 AM
Work Order: WO-01325958
Type: FCO
Priority: PC
2.0 business days
Status: Ready for Release by Mandy Newman Dec 15, 12:19 PM
From: Entered
To: Ready for Release
Duration in Previous: 3d 1h
🚨 40.0 business days
👷
2 Technician Assigned Feb 11, 12:16 PM
Technician Assigned → Status: Assigned
👷
Status: Assigned by Laura Hill
From: Ready for Release
To: Assigned
Duration in Previous: 57d 23h
3.0 business days
👷
Technician Assigned by Megan Klose Feb 13, 9:09 AM
Technician: Davon Cox
🗓️
2 Status: Scheduled Feb 13, 9:22 AM
Status: Scheduled → Dispatch Scheduled
📅
Dispatch Scheduled by Megan Klose
Scheduled For: 2026-02-16T13:00:00.000+0000
4.0 business days
⏱️
2 3x Labor Added Feb 19, 11:34 AM
3x Labor Added → Status: Completed
✔️
Status: Completed by Davon Cox
From: Scheduled
To: Completed
Duration in Previous: 6d 2h
🗓️
Status: Scheduled by Davon Cox Feb 19, 8:34 PM
From: Completed
To: Scheduled
Duration in Previous: 8h 56m
1.4 business days
✔️
Status: Completed by Davon Cox Feb 22, 5:33 PM
From: Scheduled
To: Completed
Duration in Previous: 2d 20h
🏁
Status: Closed by Davon Cox Feb 22, 5:43 PM
From: Completed
To: Closed
Duration in Previous: 10m
⚠️
Status: Incomplete by Sean Currie Feb 23, 11:28 AM
From: Closed
To: Incomplete
Duration in Previous: 17h 45m
🔩
2 Parts Added Feb 23, 11:37 AM
Parts Added → Status: Completed
✔️
Status: Completed by Davon Cox
From: Incomplete, Completed
To: Completed, Closed
Duration in Previous: 10m
⏱️
2 Labor Added Feb 24, 10:42 AM
Labor Added → Status: Ready to Bill
📄
Status: Ready to Bill by Timothy Amburgey
From: Closed
To: Ready to Bill
Duration in Previous: 23h 4m

Bottlenecks

Entered
Duration: 2.0 business days (threshold: 4.0 business hours)
Exceeded by 1.5 business days
Assigned
Duration: 2.8 business days (threshold: 4.0 business hours)
Exceeded by 2.3 business days
Scheduled
Duration: 4.0 business days (threshold: 2.0 business days)
Exceeded by 2.0 business days
Completed
Duration: 1.1 business days (threshold: 1.0 business days)
Exceeded by 0.9 business hours
Closed
Duration: 1.9 business days (threshold: 1.0 business days)
Exceeded by 7.3 business hours

🔗 Related Artifacts

No related artifacts

ℹ️ Work Order Details

Status
Ready to Bill
Priority
PC
Type
FCO
Branch
N/A
Technician
Davon Cox
Scheduled
Feb 16, 8:00 AM
Created
Dec 12, 11:06 AM
Invoice #
N/A

Work Details (5)

Name Type Status Qty Amount Date
WL-03111960 Labor Open 1.25 $100.00 Feb 19, 11:34 AM
WL-03111961 Labor Open 1 $100.00 Feb 19, 11:34 AM
WL-03111962 Labor Open 0.5 $0.00 Feb 19, 11:34 AM
WL-03116074 Parts Open 1 $0.00 Feb 23, 11:37 AM
WL-03117412 Labor Open 0.016666666666667 $38.00 Feb 24, 10:42 AM

Details