📍 PROVIDENCE MEDICAL CENTER — 8929 PARALLEL PKWY, KANSAS CITY KS, 66112
Dec 12, 2025 → Feb 27, 2026
🔴 Behind
PC
SLA Clock
Per client agreement · Target: Feb 28, 2026
✓ SLA Met
Active: 4 biz days
Paused: 47 biz days
🕐 Clock stopped: Completed (Feb 28, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸47d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Feb 24, 2026
47 business days (not counted)
▶ Clock Running
Status: Scheduled
Feb 24 → Feb 25, 2026
1 business day (counted)
Running total: 2 of ? biz days used
▶ Clock Running
Status: Tech On Site
Feb 25 → Feb 28, 2026
2 business days (counted)
Running total: 4 of ? biz days used
Created: Dec 12, 2025Completed: Feb 28, 2026
Show SLA Clock Detail (4 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 12 → Dec 15
Created
▶ Running
1
1
Dec 15 → Feb 24
Ready for Release
⏸ Paused
47
—
Feb 24 → Feb 25
Scheduled
▶ Running
1
2
Feb 25 → Feb 28
Tech On Site
▶ Running
2
4
Dispatch2.0d
Holding47.9d (excluded)
Coordinators1.9d
Field Work3.0d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
47.9 days
Coordinators
1.9 days
Field Work
3.0 days
Over SLA
Created: Dec 12, 2025
Due: Feb 28, 2026 (5 days overdue)
📅
52.0 daysTotal Age
⏱️
47.9d in Ready for ReleaseLongest Stage
🔄
6 transitionsStatus Changes
⚠️
3.5d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
3 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
On Track
Ready to Bill • 82 days old
🚨
Primary Delay47.9 days stuck in "Ready for Release"
⏱️
Total Delay3.5 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
47.9d
Scheduled
1.9d / 2.0d SLA
Tech On Site
3.0d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
1.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/24/2026 DMalloy - POCs located on a PWO for this location. Emailed over the recall deatils and advised FSE Bennet would be on site this week between tomorrow and Friday for the device remediations.
abenitez@primehealthcare.com
lkraft@primehealthcare.com
02/24/2026: ARoosa- Called; 913-596-4000 xfer to biomed opt. 5 for Luke Craft Biomed lead. VM/LM
02/16/2026 MHamersley - called 913-596-4000, xfer to Bio Med, call went to Bill Peterson's VM/LM,
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R3.4 or R3.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional”.
🔍Technician Findings
N/A
⚠️Problem Description
FMI 78101C: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📞Call Description
FMI 78101
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - PROVIDENCE MEDICAL CENTER - KANSAS CITY, KS - WO-01326002
Completed Feb 24, 2026
To: lkraft@primehealthcare.com; abenitez@primehealthcare.com
CC: central.scheduling@emsar.com; bennet.dartey@emsar.com; fmessak@emsar.com
BCC: deon.malloy@emsar.com
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - PROVIDENCE MEDICAL CENTER - KANSAS CITY, KS - WO-01326002
Body:
Hello,
Thank you for taking my call today. As discussed, I am contacting yo
📧
Email
Completed Feb 24, 2026
02/24/2026 DMalloy - POCs located on a PWO for this location. Emailed over the recall deatils adn advised FSE Bennet would be on site this week between tomorrow and Friday for the device remediations.
abenitez@primehealthcare.com
lkraft@primehealthcare.com
📞
Live Call_Left Message
Completed Feb 24, 2026
02/24/2026: ARoosa- Called; 913-596-4000 xfer to biomed opt. 5 for Luke Craft Biomed lead. VM/LM
📞
Live Call_Left Message
Completed Feb 16, 2026
02/16/2026 MHamersley - called 913-596-4000, xfer to Bio Med, call went to Bill Peterson's VM/LM,
📞
Live Call_Left Message
Completed Jan 30, 2026
01/30/2026- ARoosa Called 913-596-4000, xfer to Bio med VM/LM
📅
WO-01326002 - FCO - KANSAS CITY - PROVIDENCE MEDICAL CENTER
Feb 25, 2026 8:00am – 5:00pm
Work Order: WO-01326002
Account: GE Healthcare FMI
Customer Case Number: 1-554880127614, 1-554907756670
Location: PROVIDENCE MEDICAL CENTER
Address: 8929 PARALLEL PKWY, KANSAS CITY, KS 66112
Manufacturer:
Model #:
Serial #: VFC001373
Contact Name: Luke Kraft
Contact Phone: 913-207-9200 /