📍 TEXAS JOINT INSTITUTE MCKINNEY — 4201 MEDICAL CENTER DR STE 100, MCKINNEY TX, 75069
Dec 12, 2025 → Feb 18, 2026
🔴 Behind
PC
SLA Clock
Per client agreement · Target: Feb 28, 2026
✓ SLA Met (7 biz days early)
Active: 3 biz days
Paused: 41 biz days
🕐 Clock stopped: Completed (Feb 18, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸41d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Feb 16, 2026
41 business days (not counted)
▶ Clock Running
Status: Assigned
Feb 16 → Feb 16, 2026
0 business days (counted)
Running total: 1 of ? biz days used
▶ Clock Running
Status: Scheduled
Feb 16 → Feb 18, 2026
2 business days (counted)
Running total: 3 of ? biz days used
▶ Clock Running
Status: Tech On Site
Feb 18 → Feb 18, 2026
0 business days (counted)
Running total: 3 of ? biz days used
Created: Dec 12, 2025Completed: Feb 18, 2026
Show SLA Clock Detail (5 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 12 → Dec 15
Created
▶ Running
1
1
Dec 15 → Feb 16
Ready for Release
⏸ Paused
41
—
Feb 16 → Feb 16
Assigned
▶ Running
0
1
Feb 16 → Feb 18
Scheduled
▶ Running
2
3
Feb 18 → Feb 18
Tech On Site
▶ Running
0
3
Dispatch2.0d
Holding41.8d (excluded)
Coordinators2.0d
Field Work0.3d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Holding(excluded from total)
41.8 days
Coordinators
2.0 days
Field Work
0.3 days
Created: Dec 12, 2025
Due: Feb 28, 2026 (5 days overdue)
📅
45.0 daysTotal Age
⏱️
41.8d in Ready for ReleaseLongest Stage
🔄
6 transitionsStatus Changes
⚠️
7.7d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
5 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
At Risk - Monitor Closely
Ready to Bill • 82 days old
🚨
Primary Delay41.8 days stuck in "Ready for Release"
⏱️
Total Delay7.7 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
41.8d
Assigned
0.0d / 0.5d SLA
Scheduled
2.0d / 2.0d SLA
Tech On Site
0.3d / 1.0d SLA
Closed
7.2d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
CONFIRMED
02/16/2026 LHill-rec'd email from Joline-That works for us.
See logged email.
02/16/2026 LHill-sent email to Joline-To move forward, we can have our FSE onsite on 02/18/2026, with an arrival time between 12:30 PM-2:30PM. Please confirm that this works for your facility.
2/16/2026 LHill-rec'd call from Joline Christoffersen. She advised that anytime on Wednesday 02/18 between 8-5 and Thursday 02/19 8-10 would work. I advised I would send an email to confirm time. joline.christoffersen@hcahealthcare.com
02/16/2026 LHill-Called 469-796-4295 . Receptionist advised I need to speak with Jolene or Jenna. Left Jolene a message. (I am trying to offer 02/18 arrive between 1-3 place on calendar for 2:00. Only 3 mins away WO-01325732.)
02/05/2026 YEspino Called 469-796-4295 to verify the device in question. Was unable to speak to anybody or leave a message because once dialed all i heard was a busy signal.
🔧Work Performed
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R4.5 or R4.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”.
🔍Technician Findings
FMI 78101
⚠️Problem Description
FMI 78101E: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📞Call Description
FMI 78101
📧
Email:Re: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - TEXAS JOINT INSTITUTE MCKINNEY - MCKINNEY, TX - WO-01326152
Completed Feb 16, 2026
From: Joline.Christoffersen@hcahealthcare.com
To: laura.hill@emsar.com
Cc: central.scheduling@emsar.com, fmessak@emsar.com, rhea.berry@emsar.com
Hi Laura, That works for us. Joline Christoffersen, CPC, CPPM | Practice Manager-McKinney/Sherman/Allen, Texas Joint InstitutePhysician Services Group | HCAC: 916-599-3342Joline.Christoffersen@HCAhealthcare.com“Not all of us can do great things. But we can do small things with great love.”- Mother Teresa| This email and any files transmitted wi
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - TEXAS JOINT INSTITUTE MCKINNEY - MCKINNEY, TX - WO-01326152
Completed Feb 16, 2026
To: joline.christoffersen@hcahealthcare.com
CC: central.scheduling@emsar.com; fmessak@emsar.com; rhea.berry@emsar.com
BCC: laura.hill@emsar.com
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - TEXAS JOINT INSTITUTE MCKINNEY - MCKINNEY, TX - WO-01326152
Body:
Hi Joline,
Thank you for taking my call today. As discussed, I am contacting you on behalf of
📋
Inbound from +19725665255
Completed Feb 16, 2026
2/16/2026 LHill-rec'd call from Joline Christoffersen. She advised that anytime on Wednesday 02/18 between 8-5 and Thursday 02/19 8-10 would work. I advised I would send an email to confirm time. joline.christoffersen@hcahealthcare.com
📋
Outbound to +14697964295
Completed Feb 16, 2026
02/16/2026 LHill-Called 469-796-4295 . Receptionist advised I need to speak with Jolene or Jenna. Left Jolene a message. (I am trying to offer 02/18 arrive between 1-3 place on calendar for 2:00. Only 3 mins away WO-01325732.)
📞
Live Call_No Answer Unable to Leave Message
Completed Feb 5, 2026
02/05/2026 YEspino Called 469-796-4295 to verify the device in question. Was unable to speak to anybody or leave a message because once dialed all i heard was a busy signal.
📅
WO-01326152 - FCO - MCKINNEY - TEXAS JOINT INSTITUTE MCKINNEY
Feb 18, 2026 12:30pm – 2:30pm
Work Order: WO-01326152
Account: GE Healthcare FMI
Customer Case Number: 1-554880491136, 1-554907964032
Location: TEXAS JOINT INSTITUTE MCKINNEY
Address: 4201 MEDICAL CENTER DR STE 100, MCKINNEY, TX 75069
Manufacturer:
Model #:
Serial #: VFD000758
Contact Name: Joline. Cristoffersen
Contact Phone: