📍 TRISTAR JOINT REPLACEMENT INSTITUTE — 1001 HEALTH PARK DR FL 2, BRENTWOOD TN, 37027
Dec 12, 2025 → Feb 18, 2026
🔴 Behind
PC
SLA Clock
Per client agreement · Target: Feb 28, 2026
✓ SLA Met (7 biz days early)
Active: 4 biz days
Paused: 40 biz days
🕐 Clock stopped: Completed (Feb 18, 2026)
▶ Clock Running
Status: Created
Dec 12 → Dec 15, 2025
1 business day (counted)
Running total: 1 of ? biz days used
⏸39d
⏸ Clock Paused
Status: Ready for Release
Dec 15 → Feb 11, 2026
39 business days (not counted)
▶ Clock Running
Status: Scheduled
Feb 11 → Feb 11, 2026
0 business days (counted)
Running total: 1 of ? biz days used
⏸ Clock Paused
Status: Awaiting Parts
Feb 11 → Feb 12, 2026
1 business day (not counted)
3d
▶ Clock Running
Status: Scheduled
Feb 12 → Feb 18, 2026
3 business days (counted)
Running total: 4 of ? biz days used
▶ Clock Running
Status: Tech On Site
Feb 18 → Feb 18, 2026
0 business days (counted)
Running total: 4 of ? biz days used
Created: Dec 12, 2025Completed: Feb 18, 2026
Show SLA Clock Detail (6 periods)
Period
Status
Clock
Biz Days
Running Total
Dec 12 → Dec 15
Created
▶ Running
1
1
Dec 15 → Feb 11
Ready for Release
⏸ Paused
39
—
Feb 11 → Feb 11
Scheduled
▶ Running
0
1
Feb 11 → Feb 12
Awaiting Parts
⏸ Paused
1
—
Feb 12 → Feb 18
Scheduled
▶ Running
3
4
Feb 18 → Feb 18
Tech On Site
▶ Running
0
4
Dispatch2.0d
Material Management1.4d
Holding39.8d (excluded)
Coordinators3.6d
Field Work0.3d
Billing3.0d
✓ SLA Met
Dispatch
2.0 days
Over SLA
Material Management
1.4 days
Holding(excluded from total)
39.8 days
Coordinators
3.6 days
Over SLA
Field Work
0.3 days
Billing
3.0 days
Over SLA
Created: Dec 12, 2025
Due: Feb 28, 2026 (5 days overdue)
📅
45.0 daysTotal Age
⏱️
39.8d in Ready for ReleaseLongest Stage
🔄
9 transitionsStatus Changes
⚠️
8.5d overTotal Delay
📦
1 orders, 0 quotesParts & Quotes
📝
7 linesWork Details
🏠
1 visitOn-Site Trips
👤
Self-ScheduledScheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
At Risk - Monitor Closely
Ready to Bill • 82 days old
🚨
Primary Delay39.8 days stuck in "Ready for Release"
⏱️
Total Delay8.5 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
2.0d / 0.5d SLA
Ready for Release
39.8d
Scheduled
3.6d / 2.0d SLA
Awaiting Parts
1.4d / 3.0d SLA
Tech On Site
0.3d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
4.5d / 1.0d SLA
Ready to Bill
3.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/11/2026 RArnold - Called 615-416-6453 for Siobhan Parks. She confrimed device in another email. She needs a wednesday. Dispatching to Ryan for 02/18. ATTN: Siobhan Parks
Siobhan.Parks@hcahealthcare.com
01/22/2026 MHamersley/ARoosa - called 615-342-0038, spoke to Peggy Martin, Office Mgr, sent email: peggy.martin@hcahealthcare.com
🔧Work Performed
"Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing SW Version R3.4 or R3.5 and communicated with customer to scrap previous software versions”.
Verification Test: “Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed. Completed full system diagnostics-system passed.
Equipment passed all required tests and has been returned to the customer fully
functional”.
Template Name:
Template Description:
Problem Found:
Action Taken: “Completed FMI 78101 by installing new supplied batteries / Batteries installed by
customer and dispose of old batteries according to local regulation”.
Verification Test: “Performed battery charging test - verified after 10 minutes, the scan time increased. Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully functional”."
🔍Technician Findings
FMI
⚠️Problem Description
FMI 78101B: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R3.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📞Call Description
FMI 78101
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - TRISTAR JOINT REPLACEMENT INSTITUTE - BRENTWOOD, TN - WO-01326163
Completed Feb 11, 2026
To: Siobhan.Parks@hcahealthcare.com; ryan.lewis@emsar.com
CC: central.scheduling@emsar.com; fmessak@emsar.com; douglas.pearce@emsar.com; brandi.rodriguez@emsar.com; rebecca.arnold@emsar.com
BCC: rebecca.arnold@emsar.com
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - TRISTAR JOINT REPLACEMENT INSTITUTE - BRENTWOOD, TN - WO-01326163
Body:
Hi Siobhan,
📧
Live Call_Followed Up With Email
Completed Jan 22, 2026
01/22/2026 ARoosa- 01/22/2026 MHamersley/ARoosa - called 615-342-0038, spoke to Peggy Martin, Office Mgr, sent email: peggy.martin@hcahealthcare.com:
📧
Email: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - TRISTAR JOINT REPLACEMENT INSTITUTE - BRENTWOOD, TN - WO-01326163
Completed Jan 22, 2026
To: peggy.martin@hcahealthcare.com
CC: angela.roosa@emsar.com; mary.hamersley@emsar.com; brandi.rodriguez@emsar.com; fmessak@emsar.com
BCC:
Attachment: Vendor Letter for Customers 78101 EMSAR.pdf, FMI 78101 Customer Letter_DOC3115777_Rev2.pdf
Subject: Field Action Notice - FMI 78101 - GE Healthcare - Venue Go/Fit Thermal Event - TRISTAR JOINT REPLACEMENT INSTITUTE - BRENTWOOD, TN - WO-01326163
Body:
Good Afternoon Peggy,
Thank you for taking my call today. As discussed, I am contacting you
📧
Live Call_Followed Up With Email
Completed Jan 22, 2026
01/22/2026 MHamersley/ARoosa - called 615-342-0038, spoke to Peggy Martin, Office Mgr, sent email: peggy.martin@hcahealthcare.com
📅
WO-01326163 - FCO - BRENTWOOD - TRISTAR JOINT REPLACEMENT INSTITUTE
Feb 18, 2026 8:00am – 5:00pm
Work Order: WO-01326163
Account: GE Healthcare FMI
Customer Case Number: 1-554879921338, 1-554907719778
Location: TRISTAR JOINT REPLACEMENT INSTITUTE
Address: 1001 HEALTH PARK DR FL 2, BRENTWOOD, TN 37027
Manufacturer:
Model #:
Serial #: VGC006760
Contact Name: Siobhan Parks
Contact Phone: 615-416
JPG
VGC006760_SN
Feb 18, 2026
JPG
VGC006760_Batteries
Feb 18, 2026
JPG
VGC006760_Software
Feb 18, 2026
Timeline
📋
Work Order Created Dec 12, 11:07 AM
Work Order:WO-01326163
Type:FCO
Priority:PC
⏳2.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:14 PM
From:Entered
To:Ready for Release
Duration in Previous:3d 1h
🚨40.0 business days
👷
▶
5
Technician Assigned Feb 11, 3:40 PM
Technician Assigned → Status: Scheduled → Dispatch Scheduled → Parts Requested → Status: Awaiting Parts
🗓️
Status: Scheduled by Rebecca Arnold
From:Ready for Release
To:Scheduled
Duration in Previous:58d 3h
📅
Dispatch Scheduled by Rebecca Arnold
Scheduled For:2026-02-18T13:00:00.000+0000
📦
Parts Requested
Order:00090195
Status:Closed
📦
Status: Awaiting Parts by Rebecca Arnold
From:Scheduled
To:Awaiting Parts
Duration in Previous:2m
🔩
▶
2
2x Parts Added Feb 12, 2:24 PM
2x Parts Added → Status: Scheduled
🗓️
Status: Scheduled by Sean Weaver
From:Awaiting Parts
To:Scheduled
Duration in Previous:22h 41m
⏳4.0 business days
🔧
▶
3
Status: Tech On Site Feb 18, 1:08 PM
Status: Tech On Site → Labor Added → 2x Travel Added
⏱️
Labor Added
Type:Labor
Status:Open
Qty:1.25
🚗
2x Travel Added
Type:Travel
Status:Open
Qty:63, 12
⏱️
▶
3
Labor Added Feb 18, 3:16 PM
Labor Added → Status: Completed → Status: Closed
✔️
Status: Completed by Ryan Lewis
From:Tech On Site
To:Completed
Duration in Previous:2h 7m
🏁
Status: Closed by Ryan Lewis
From:Completed
To:Closed
Duration in Previous:4m
⚠️5.0 business days
⏱️
▶
2
Labor Added Feb 24, 11:08 AM
Labor Added → Status: Ready to Bill
📄
Status: Ready to Bill by Timothy Amburgey
From:Closed
To:Ready to Bill
Duration in Previous:5d 19h
⏳3.0 business days
💵
Status: Billing Review by Timothy AmburgeyFeb 26, 11:26 AM
From:Ready to Bill, Billing Review
To:Billing Review, Ready to Bill
Duration in Previous:2d 0h, 0m
Bottlenecks
Entered
Duration: 2.0 business days (threshold: 4.0 business hours)
Exceeded by 1.5 business days
Scheduled
Duration: 3.6 business days (threshold: 2.0 business days)
Exceeded by 1.6 business days
Closed
Duration: 4.5 business days (threshold: 1.0 business days)
Exceeded by 3.5 business days
Ready to Bill
Duration: 3.0 business days (threshold: 1.0 business days)