📍 MEDINA HOSPITAL — 1000 E WASHINGTON ST, MEDINA OH, 44256
Jan 6, 2026 → Jan 14, 2026
✅ Completed
PC
SLA Clock
Per client agreement · Target: Jan 31, 2026
✓ SLA Met (11 biz days early)
Active: 6 biz days
🕐 Clock stopped: Completed (Jan 14, 2026)
▶ Clock Running
Status: Created
Jan 6 → Jan 6, 2026
0 business days (counted)
Running total: 0 of ? biz days used
5d
▶ Clock Running
Status: Ready for Scheduling
Jan 6 → Jan 13, 2026
5 business days (counted)
Running total: 5 of ? biz days used
1d
▶ Clock Running
Status: Scheduled
Jan 13 → Jan 14, 2026
1 business day (counted)
Running total: 6 of ? biz days used
▶ Clock Running
Status: Tech On Site
Jan 14 → Jan 14, 2026
0 business days (counted)
Running total: 6 of ? biz days used
Created: Jan 6, 2026Completed: Jan 14, 2026
Show SLA Clock Detail (4 periods)
Period
Status
Clock
Biz Days
Running Total
Jan 6 → Jan 6
Created
▶ Running
0
0
Jan 6 → Jan 13
Ready for Scheduling
▶ Running
5
5
Jan 13 → Jan 14
Scheduled
▶ Running
1
6
Jan 14 → Jan 14
Tech On Site
▶ Running
0
6
Dispatch0.0d
Coordinators7.3d
Field Work1.5d
Billing4.7d
✓ SLA Met
Dispatch
0.0 days
Coordinators
7.3 days
Over SLA
Field Work
1.5 days
Over SLA
Billing
4.7 days
Over SLA
Created: Jan 6, 2026
Due: Jan 31, 2026
Completed: Jan 15, 2026 (16 days early)
📅
6.6 daysTotal Age
⏱️
11.0d in ClosedLongest Stage
🔄
8 transitionsStatus Changes
⚠️
16.9d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
9 linesWork Details
🏠
2 visitsOn-Site Trips
👤
Self-ScheduledScheduling
🔄
2 On-Site Visits
2 on-site visits — follow-up trip required
👤 Self-scheduled by Samuel Biddie — scheduling delays attributed to technician, not coordinators
Visit 1Jan 13, 2026
Visit 2Jan 14, 2026
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with significant delays
Invoiced • 57 days old
🚨
Primary Delay11.0 days stuck in "Closed"
⏱️
Total Delay16.9 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
5.3d / 1.0d SLA
Scheduled
2.0d / 2.0d SLA
Tech On Site
0.5d / 1.0d SLA
Completed
1.1d / 1.0d SLA
Closed
11.0d / 1.0d SLA
Ready to Bill
3.5d / 1.0d SLA
Invoice Pending
1.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
01/06/2026 ABelyea - device was previously completed prior to maintenance software release. GE requested we return to upgrade to the newest software as customer is having issues with the device. No battery needed.
🔧Work Performed
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R3.4 and
communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed
all probes pass. Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed. Equipment passed
all required tests and has been returned to the customer fully
functional”.
🔍Technician Findings
FMI needed
⚠️Problem Description
FMI 78101E: Venue Fit Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Work Order: WO-01333009
Account: GE Healthcare FMI
Customer Case Number: 1-556928316231
Location: MEDINA HOSPITAL
Address: 1000 E WASHINGTON ST, MEDINA, OH 44256
Manufacturer: GE Healthcare
Model #: Venue Fit Ultrasound
Serial #: RVFD000952
Contact Name: Abigail Gray
Contact Phone: (256) 426-2365
Co