📍 BON SECOURS ST FRANCIS MEDICAL CENTER - Midlothian — 13710 St Francis Blvd, Midlothian VA, 23114
Feb 19, 2026 → Feb 20, 2026
✅ Completed
P5
SLA Clock
5 business days from creation · Target: Feb 26, 2026
✓ SLA Met (4 biz days early)
Active: 1 biz days
Budget: 5 biz days
🕐 Clock stopped: Completed (Feb 21, 2026)
SLA Target
▶ Clock Running
Status: Created
Feb 19 → Feb 19, 2026
0 business days (counted)
Running total: 0 of 5 biz days used
▶ Clock Running
Status: Ready for Scheduling
Feb 19 → Feb 19, 2026
0 business days (counted)
Running total: 0 of 5 biz days used
1d
▶ Clock Running
Status: Completed
Feb 19 → Feb 20, 2026
1 business day (counted)
Running total: 1 of 5 biz days used
0d
▶ Clock Running
Status: Assigned
Feb 20 → Feb 21, 2026
0 business days (counted)
Running total: 1 of 5 biz days used
Created: Feb 19, 2026Completed: Feb 21, 2026
Business Days Used1 / 5
05d budget
Show SLA Clock Detail (4 periods)
Period
Status
Clock
Biz Days
Running Total
Feb 19 → Feb 19
Created
▶ Running
0
0 / 5
Feb 19 → Feb 19
Ready for Scheduling
▶ Running
0
0 / 5
Feb 19 → Feb 20
Completed
▶ Running
1
1 / 5
Feb 20 → Feb 21
Assigned
▶ Running
0
1 / 5
Dispatch0.0d
Coordinators1.1d
Field Work1.1d
Billing0.0d
✓ SLA Met
Dispatch
0.0 days
Coordinators
1.1 days
Over SLA
Field Work
1.1 days
Over SLA
Billing
0.0 days
Created: Feb 19, 2026
Due: Feb 26, 2026
Completed: Feb 21, 2026 (5 days early)
📅
1.3 daysTotal Age
⏱️
7.0d in ClosedLongest Stage
🔄
7 transitionsStatus Changes
⚠️
6.6d overTotal Delay
📦
0 orders, 0 quotesParts & Quotes
📝
10 linesWork Details
🏠
2 visitsOn-Site Trips
📋
CoordinatorScheduling
🔄
2 On-Site Visits
2 on-site visits — follow-up trip required
Visit 1Feb 19, 2026
Visit 2Feb 20, 2026
AI Analysis
Powered by OpenAI
Generating AI analysis...
Completed with minor delays
Invoiced • 13 days old
🚨
Primary Delay7.0 days stuck in "Closed"
⏱️
Total Delay6.6 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
0.2d / 1.0d SLA
Completed
1.1d / 1.0d SLA
Assigned
1.0d / 0.5d SLA
Closed
7.0d / 1.0d SLA
Ready to Bill
0.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
2-20 Mburris added tech per tech supervisor
🔧Work Performed
Tested bed in multiple positions, tested all functions, confirmed no faults in the system. The nurses most likely used the accessory power cord to plug into the wall instead of the bed main power cord.
🔍Technician Findings
When plugged in using the correct cord, no faults or errors were detected, side screens are functional and all features work.
⚠️Problem Description
Bed side rails won't turn on or load
Bed repair
Lemos, Fernando
Timeline
📋
▶
2
Work Order Created Feb 19, 4:38 PM
Work Order Created → Status: Ready for Scheduling
📅
Status: Ready for Scheduling by Sherry Reich
From:Entered
To:Ready for Scheduling
Duration in Previous:0m
✔️
Status: Completed by Fernando LemosFeb 19, 5:51 PM