Work Order 360 View
Detailed work order timeline, KPIs, and analysis
PC
SLA Clock
Per client agreement · Target: Dec 31, 2026
✓ SLA Met (298 biz days early)
🕐 Clock stopped: Completed (Oct 18, 2025)
▶ Clock Running
Status: Created
Jul 15 → Jul 15, 2025
0 business days (counted)
Running total: 0 of ? biz days used
▶ Clock Running
Status: Scheduled
Jul 15 → Jul 21, 2025
4 business days (counted)
Running total: 4 of ? biz days used
59d
▶ Clock Running
Status: Assigned
Jul 21 → Oct 14, 2025
59 business days (counted)
Running total: 63 of ? biz days used
▶ Clock Running
Status: Scheduled
Oct 14 → Oct 14, 2025
0 business days (counted)
Running total: 63 of ? biz days used
▶ Clock Running
Status: Tech On Site
Oct 14 → Oct 18, 2025
3 business days (counted)
Running total: 66 of ? biz days used
Created: Jul 15, 2025
Completed: Oct 18, 2025
Show SLA Clock Detail (5 periods)
Period
Status
Clock
Biz Days
Running Total
Jul 15 → Jul 15
Created
▶ Running
0
0
Jul 15 → Jul 21
Scheduled
▶ Running
4
4
Jul 21 → Oct 14
Assigned
▶ Running
59
63
Oct 14 → Oct 14
Scheduled
▶ Running
0
63
Oct 14 → Oct 18
Tech On Site
▶ Running
3
66
Coordinators
64.7 days
Over SLA
Field Work
3.7 days
Over SLA
Created: Jul 15, 2025
Due: Dec 31, 2026
Completed: Oct 18, 2025 (439 days early)
⏱️
60.0d in Assigned
Longest Stage
🔄
7 transitions
Status Changes
⚠️
64.9d over
Total Delay
📦
0 orders, 0 quotes
Parts & Quotes
👤
Self-Scheduled
Scheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
🎫
Case #00510573
URGENT Philips WO Request
Closed
Trilogy EV300 Ventilator - Machine Flow Sensor Inaccurate Flow Measurements and/or Delivery
🚨
Primary Delay
60.0 days stuck in "Assigned"
⏱️
Total Delay
64.9 business days over SLA thresholds
SLA vs Actual Time by Status
Scheduled
4.7d
/ 2.0d SLA
Assigned
60.0d
/ 0.5d SLA
Tech On Site
3.7d
/ 1.0d SLA
Completed
0.0d
/ 1.0d SLA
💬 Activities, Notes & Communications
📝 WO Notes (6)
📋 Tasks (1)
💬 Chatter (1)
📅 Events (1)
🔒
Internal CSA Notes
10/14/2025 EMartin dispatching to FSE onsite per request
07/15/2025 EMartin dispatching to FSE onsite per request for SN# H2812309845AA
located at UNC Lenoir Health Care, Kinston, NC
🔧
Work Performed
10/14 - while onsite, attempted to perform FCO86600081 per instruction. 2 devices failed preliminary Tests.
H288179901FAF - AECM Verification: Reading 29.08 cmH2O LSL- 25.50 USL- 28.50 FAIL
H2812309845AA - AECM Verification: Reading 30.01 cmH2O LSL- 25.50 USL- 28.50 FAIL
Incident WOs created, (PWO-13722271 & PWO-13723104) quotes requested.
🔍
Technician Findings
FCO86600081
⚠️
Problem Description
Trilogy EV300 Ventilator - Machine Flow Sensor Inaccurate Flow Measurements and/or Delivery
📄
Description
Philips Trilogy EV300 Hospital Ventilator (DS2200X11B / 989805661881)
📞
Call Description
FCO86600081
📋
0-0-0
Completed
Nov 12, 2025
Tech Supervisor Spot Check
2. # of Critical Defects: 0
3. # of Major Defects: 0
4. # of Minor Defects: 0
5. # of EMSAR Defects: 0
6. FCO 81 EV300
7. i. Tools: Certifier, multimeter, electrical safety analyzer,) - YES
8. ii. Prontoforms – uploaded and correct template - YES
9. iii. Event Log – uploaded - YES
10. iv. CSR – signed and uploaded to ServiceMax - YES
11. v. EMSAR Requirements (do not include in task subject count): - Picture of device- YES
FOLLOW UP WORK REQUIRED
Jul 16, 2025 4:03pm
@Jessica Foley Missing Required Parts List
📅
WO-01292006 - FCO - Trilogy EV300 Vent - Kinston - UNC Lenoir Health Care, Kinston, NC
Oct 13, 2025 8:00pm – 8:00pm
Work Order: WO-01292006
Account: PHILIPS HEALTHCARE
Customer Case Number: WO-13166162
Location: UNC Lenoir Health Care, Kinston, NC
Address: 100 Airport Rd, Kinston, NC 28501
Manufacturer: Philips Healthcare
Model #: Trilogy EV300 Vent
Serial #: H283493166210
Contact Name: Paul Silberman
Contact Pho
Timeline
📞
Case Number:
00510573
Subject:
URGENT Philips WO Request
Status:
Closed
📋
Work Order:
WO-01292006
Type:
FCO
Priority:
PC
👷
📅
Scheduled For:
2025-07-15T04:00:00.000+0000, 2025-07-15T00:00:00.000+0000, 2025-07-15T12:00:00.000+0000
🗓️
From:
Entered
To:
Scheduled
Duration in Previous:
3h 8m
👷
From:
Scheduled
To:
Assigned
Duration in Previous:
5d 18h
👷
🗓️
From:
Assigned
To:
Scheduled
Duration in Previous:
85d 5h
📅
Scheduled For:
2025-10-14T16:00:00.000+0000, 2025-10-14T12:00:00.000+0000
🔧
From:
Scheduled
To:
Tech On Site
Duration in Previous:
10m
⏱️
Type:
Labor
Status:
Open
Qty:
1.5, 2
✔️
From:
Tech On Site
To:
Completed
Duration in Previous:
3d 22h
🏁
From:
Completed
To:
Closed
Duration in Previous:
2h 46m
📄
From:
Closed, Ready to Bill, Invoice Pending
To:
Ready to Bill, Invoice Pending, Invoiced
Duration in Previous:
1d 18h, 0m
🧾
Invoice Number:
SM-244123
Bottlenecks
Scheduled
Duration: 4.7 business days (threshold: 2.0 business days)
Exceeded by 2.7 business days
Assigned
Duration: 60.0 business days (threshold: 4.0 business hours)
Exceeded by 59.5 business days
Tech On Site
Duration: 3.7 business days (threshold: 1.0 business days)
Exceeded by 2.7 business days
ℹ️ Work Order Details
Technician
Patricia Gordon
Scheduled
Oct 14, 8:00 AM
Work Details (2)
Name
Type
Status
Qty
Amount
Date
WL-02971356
Labor
Open
1.5
$0.00
Oct 15, 12:15 PM
WL-02971357
Labor
Open
2
$0.00
Oct 15, 12:15 PM