Device failed during SYSTEM SETUP TEST, Active Exhalation Verification, Pilot: Reading 29.88 cmH2O LSL 25.50 USL 28.50 FAIL. FCO86600081: Follow-up Repair to be Handled by Philips ASP EMSAR.
09/22/2025 SRobinson- WO-01305391 already created for this request.
08/18/2025 Emartin sent email to Caleb FSE will be onsite Mon 08/25/25 parts show delivered
08/12/2025 Emartin sent email to Caleb advising FSE onsite Thurs 08/21/25 to complete devices to allow time for parts to arrive 08/11/2025 Emartin sent email to Caleb advising FSE onsite friday 08/15/25 to complete devices
07/28/2025 EMartin pending entitlement FSE working with POC to get hard copy PO and quotes
07/23/2025 EMartin pending entitlement FSE to work with POC to get hard copy PO and quotes updated
🔧Work Performed
Travel to site. Performed repair - replaced 3-way Solenoid & Proportional Valve, Final Test failed, blower will not start. Blower Calibration failed. created follow up activity (PWO-13577438) for return visit.
🔍Technician Findings
AECM failure, repair required
⚠️Problem Description
Device failed during SYSTEM SETUP TEST, Active Exhalation Verification, Pilot: Reading 29.88 cmH2O LSL 25.50 USL 28.50 FAIL. FCO86600081: Follow-up Repair to be Handled by Philips ASP EMSAR.