38R
SYS-LC-1260IIE
Contract ends 11/29/2025 includes Repair
PMPLUS
PM Plus - allows a lamp to be ordered and consumed on a PM if needed. Sometimes a lamp may have been installed recently, so a new lamp is not needed at time of PM. The tech needs to ask the customer these questions when scheduling the PM for a PM PLUS. PM PLUS = UV lamp is included in the service.
G7114A
DEACX26147 1260 Infinity II VW Detector
G7129A
DEAEQ47009 1260 Infinity II Vialsampler
G7110B
DEAEH01938 1260 Infinity II Isocratic Pump
G7116A
DEAEM09350 1260 Infinity II Multicolumn Thermostat
G7162A
DEAC907043 1260 Infinity II RI detector
10/3/2025 Sramen - found this WO created with the wrong SMC. I corrected to Agilent ASP
09/12 an email was sent to lsca dispatch because due date and due date description . MD
Please assign to Maria Davis ARP to site is 110 miles. Zone 3
🔧Work Performed
12/02/2025 LC (VWD/RID) PM was completed per Agilent. Solvent is dangerous THF and column were bypass to run tests with H2O. All tests passed. After system was flushed with THF before connecting the columns back. After RID exchanged and Previous PM customer stated that the RID standard injection was not as expected. After PM an injection was made (45 min) base line at the end got higher, RID will be flushed over night and customer will run blanks and calibrate.
⚠️Problem Description
38R
SYS-LC-1260IIE
Contract ends 11/29/2025 includes Repair
PMPLUS
PM Plus - allows a lamp to be ordered and consumed on a PM if needed. Sometimes a lamp may have been installed recently, so a new lamp is not needed at time of PM. The tech needs to ask the customer these questions when scheduling the PM for a PM PLUS. PM PLUS = UV lamp is included in the service.
G7114A
DEACX26147 1260 Infinity II VW Detector
G7129A
DEAEQ47009 1260 Infinity II Vialsampler
G7110B
DEAEH01938 1260 Infinity II Isocratic Pump
G7116A
DEAEM09350 1260 Infinity II Multicolumn Thermostat
G7162A
DEAC907043 1260 Infinity II RI detector
📄Description
1260 Infinity II Isocratic Pump
📞Call Description
NOV 2025 LC PMPLUS
Nov 19, 2025 1:34pm
@Maria Davis This is past scheduled over a week without any notes. What is the status?
📅
WO-01304197 - Preventive Maintenance - G7110B - Longview - Eastman Chemical Company
Dec 2, 2025 10:00am – 3:00pm
Work Order: WO-01304197
Account: AGILENT TECHNOLOGIES (ASP)
Customer Case Number: 6008007499
Location: Eastman Chemical Company
Address: 2290 Callahan Rd, Longview, TX 75602
Manufacturer:
Model #: G7110B
Serial #: DEAEH01938
Contact Name: Bennett Novak
Contact Phone: 4302401267
Contact Email: benne
JPG
1000165943
Nov 20, 2025
Timeline
📞
▶
4
Case Linked Sep 12, 2:57 PM
Case Linked → Work Order Created → Status: Ready for Scheduling → Technician Assigned
📋
Work Order Created
Work Order:WO-01304197
Type:Preventive Maintenance
Priority:P4
📅
Status: Ready for Scheduling by Leona Coonrod
From:Entered, Ready for Scheduling
To:Ready for Scheduling, Assigned
Duration in Previous:0m
👷
Technician Assigned by Leona Coonrod
Technician:a0hF00000028i6gIAA, Maria Davis
🚨21.0 business days
🗓️
▶
2
Status: Scheduled Oct 10, 12:13 PM
Status: Scheduled → Dispatch Scheduled
📅
Dispatch Scheduled by Maria Davis
Scheduled For:2025-11-13T16:00:00.000+0000
🚨28.0 business days
🔄
Status: Reschedule by Maria DavisNov 20, 8:49 PM
From:Scheduled
To:Reschedule
Duration in Previous:41d 9h
⏳3.4 business days
🗓️
▶
2
Status: Scheduled Nov 25, 8:56 AM
Status: Scheduled → Dispatch Scheduled
📅
Dispatch Scheduled by Maria Davis
Scheduled For:2025-12-02T15:00:00.000+0000
⏳4.0 business days
🔧
Status: Tech On Site by Maria DavisDec 2, 12:10 PM
From:Scheduled
To:Tech On Site
Duration in Previous:7d 3h
⏱️
▶
2
4x Labor Added Dec 2, 11:04 PM
4x Labor Added → 4x Parts Added
🔩
4x Parts Added
Type:Parts
Status:Open
Qty:1
⏳2.1 business days
🚗
Status: Tech Off Site by Maria DavisDec 4, 8:48 AM
From:Tech On Site
To:Tech Off Site
Duration in Previous:1d 20h
⏳2.0 business days
🚗
▶
2
2x Travel Added Dec 6, 5:43 PM
2x Travel Added → Status: Completed
✔️
Status: Completed by Maria Davis
From:Tech Off Site
To:Completed
Duration in Previous:2d 8h
🏁
Status: Closed by Maria DavisDec 6, 5:49 PM
From:Completed
To:Closed
Duration in Previous:4m
⏳1.0 business days
📄
Status: Ready to Bill by Hannah ShawDec 8, 3:51 PM
From:Closed
To:Ready to Bill
Duration in Previous:1d 22h
🚨11.0 business days
📨
▶
2
Status: Invoice Pending Dec 22, 1:38 PM
Status: Invoice Pending → Status: Invoiced
💰
Status: Invoiced by Hannah Shaw
From:Invoice Pending
To:Invoiced
Duration in Previous:3m
🚨14.0 business days
🧾
Invoice Created Jan 12, 3:14 PM
Invoice Number:SM-249754
Bottlenecks
Assigned
Duration: 20.6 business days (threshold: 4.0 business hours)
Exceeded by 20.1 business days
Scheduled
Duration: 28.2 business days (threshold: 2.0 business days)
Exceeded by 26.2 business days
Reschedule
Duration: 3.0 business days (threshold: 2.0 business days)
Exceeded by 1.0 business days
Scheduled
Duration: 4.0 business days (threshold: 2.0 business days)
Exceeded by 2.0 business days
Tech On Site
Duration: 2.9 business days (threshold: 1.0 business days)
Exceeded by 1.9 business days
Ready to Bill
Duration: 10.4 business days (threshold: 1.0 business days)