Work Order 360 View
Detailed work order timeline, KPIs, and analysis
PC
SLA Clock
Per client agreement · Target: Oct 3, 2025
✗ SLA Missed (6 biz days late)
🕐 Clock stopped: Completed (Oct 15, 2025)
▶ Clock Running
Status: Created
Sep 19 → Sep 19, 2025
0 business days (counted)
Running total: 0 of ? biz days used
⏸1d
⏸ Clock Paused
Status: Ready for Release
Sep 19 → Sep 22, 2025
1 business day (not counted)
5d
▶ Clock Running
Status: Assigned
Sep 22 → Sep 29, 2025
5 business days (counted)
Running total: 5 of ? biz days used
10d
▶ Clock Running
Status: Scheduled
Sep 29 → Oct 14, 2025
10 business days (counted)
Running total: 15 of ? biz days used
▶ Clock Running
Status: Tech On Site
Oct 14 → Oct 15, 2025
1 business day (counted)
Running total: 16 of ? biz days used
Created: Sep 19, 2025
Completed: Oct 15, 2025
Show SLA Clock Detail (5 periods)
| Period |
Status |
Clock |
Biz Days |
Running Total |
| Sep 19 → Sep 19 |
Created |
▶ Running
|
0 |
0 |
| Sep 19 → Sep 22 |
Ready for Release |
⏸ Paused
|
1 |
—
|
| Sep 22 → Sep 29 |
Assigned |
▶ Running
|
5 |
5 |
| Sep 29 → Oct 14 |
Scheduled |
▶ Running
|
10 |
15 |
| Oct 14 → Oct 15 |
Tech On Site |
▶ Running
|
1 |
16 |
Holding (excluded from total)
1.4 days
Coordinators
16.3 days
Over SLA
Field Work
1.9 days
Over SLA
Billing
1.4 days
Over SLA
Created: Sep 19, 2025
Due: Oct 3, 2025
Completed: Oct 15, 2025 (11 days late)
⏱️
10.6d in Scheduled
Longest Stage
🔄
9 transitions
Status Changes
⚠️
15.0d over
Total Delay
📦
0 orders, 0 quotes
Parts & Quotes
👤
Self-Scheduled
Scheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
🎫
Case #00528962
URGENT Philips WO Request
Closed
Taqiyy Coley Post FCO implementation, Device failed during SYSTEM SETUP TEST, Active Exhalation Verification, Pilot: Reading 29.88 cmH2O LSL 25.50 USL 28.50 FAIL.
🚨
Primary Delay
10.6 days stuck in "Scheduled"
⏱️
Total Delay
15.0 business days over SLA thresholds
SLA vs Actual Time by Status
Scheduled
10.6d
/ 2.0d SLA
Tech On Site
1.8d
/ 1.0d SLA
Completed
0.0d
/ 1.0d SLA
Ready to Bill
1.4d
/ 1.0d SLA
💬 Activities, Notes & Communications
🔒
Internal CSA Notes
10/13/2025 EMartin 10132025_BOUPDATE_
10/06/2025 EMartin 10062025_BOUPDATE_
09/29/2025 EMartin sent email to caleb.duranceau@trimedx.com advising FSE onsite 10/14/2025 to complete repair and remaining devices
09/22/2025 EMartin FSE sent email to quotes to get sent to Caleb Fady added Caleb to the thread for visibility
09/19/2025 EMartin FSE onsite will work with POC to get quotes and hard copy PO
🔧
Work Performed
Travel to site, get to Biomed, setup equipment. Performed Diagnostics, reloaded software, ran blower calibration, ran full Final Tests successfully.
🔍
Technician Findings
Repair required. Device had failed, blower would not start.
⚠️
Problem Description
Taqiyy Coley Post FCO implementation, Device failed during SYSTEM SETUP TEST, Active Exhalation Verification, Pilot: Reading 29.88 cmH2O LSL 25.50 USL 28.50 FAIL.
📄
Description
Philips Trilogy EV300 Hospital Ventilator (DS2200X11B / 989805661881)
📞
Call Description
Repair - Trade
📅
WO-01305391 - Repair - Trilogy EV300 Vent - Kinston - UNC Lenoir Health Care, Kinston, NC
Oct 14, 2025 8:00am – 5:00pm
Work Order: WO-01305391
Account: PHILIPS HEALTHCARE
Customer Case Number: WO-13577438
Location: UNC Lenoir Health Care, Kinston, NC
Address: 100 Airport Rd, Kinston, NC 28501
Manufacturer: Philips Healthcare
Model #: Trilogy EV300 Vent
Serial #: H28122713A507
Contact Name: Caleb Duranceau
Contact Ph
Timeline
📞
Case Number:
00528962
Subject:
URGENT Philips WO Request
Status:
Closed
📋
Work Order:
WO-01305391
Type:
Repair
Priority:
PC
✅
From:
Entered
To:
Ready for Release
Duration in Previous:
1h 44m
👷
👷
From:
Ready for Release
To:
Assigned
Duration in Previous:
2d 22h
🗓️
📅
Scheduled For:
2025-10-14T12:00:00.000+0000
🔧
From:
Scheduled
To:
Tech On Site
Duration in Previous:
14d 22h
⏱️
✔️
From:
Tech On Site
To:
Completed
Duration in Previous:
19h 42m
🏁
From:
Completed
To:
Closed
Duration in Previous:
21m
⏱️
Type:
Labor
Status:
Open
Qty:
0.016666666666667
📋
From:
Closed
To:
07e-Partial Project
Duration in Previous:
1h 21m
📄
From:
07e-Partial Project
To:
Ready to Bill
Duration in Previous:
1d 6h
💰
From:
Ready to Bill
To:
Invoiced
Duration in Previous:
10h 52m
🧾
Invoice Number:
SM-243927
Bottlenecks
Assigned
Duration: 5.6 business days (threshold: 4.0 business hours)
Exceeded by 5.1 business days
Scheduled
Duration: 10.6 business days (threshold: 2.0 business days)
Exceeded by 8.6 business days
Tech On Site
Duration: 1.8 business days (threshold: 1.0 business days)
Exceeded by 6.7 business hours
Ready to Bill
Duration: 1.4 business days (threshold: 1.0 business days)
Exceeded by 2.8 business hours
ℹ️ Work Order Details
Technician
Patricia Gordon
Scheduled
Oct 14, 8:00 AM
Work Details (4)
| Name |
Type |
Status |
Qty |
Amount |
Date |
| WL-02971119 |
Labor |
Open |
1 |
$155.00 |
Oct 15, 9:01 AM |
| WL-02971120 |
Labor |
Open |
3 |
$285.00 |
Oct 15, 9:01 AM |
| WL-02971121 |
Travel |
Open |
6 |
$0.00 |
Oct 15, 9:01 AM |
| WL-02971230 |
Labor |
Open |
0.016666666666667 |
$38.00 |
Oct 15, 10:34 AM |