***PART RETURN IS REQUIRED IF RMA IS NOT IN BOX, Email ferno@emsar.com for label***
SN 24C211160
STRETCHER LOWERS SLOWLY WITHOUT WEIGHT. CHECK MANUAL RELEASE FIRST. MAKE SURE IT ISN'T TOO TIGHT.IF THAT
DOESN'T WORK CHANGE ACTUATOR. CALL BRETT MORRIS AT FERNO FOR HELP 937.283.3605 RETURN ALL PARTS
Part ID 0806111
Return Parts: Yes
KIT, PX1 FERNO ACTUATOR
SN 24C204146
WHEEL CAME OFF. REPLACE WITH NEW WHEEL AND SEND PARTS BACK. RETURN ALL PARTS
Part ID 0806069
Return Parts: Yes
KIT, WHEEL FORK LOAD END 6X2
SN 21S023590
STRETCHER LOWERS SLOWLY WITHOUT WEIGHT. CHECK MANUAL RELEASE FIRST. MAKE SURE IT ISN'T TOO TIGHT.IF THAT
DOESN'T WORK CHANGE ACTUATOR. CALL BRETT MORRIS AT FERNO FOR HELP 937.283.3605. RETURN ALL PARTS
Part ID 0806111
Return Parts: Yes
KIT, PX1 FERNO ACTUATOR
Primary Delay8.0 days stuck in "Awaiting Parts - Customer"
⏱️
Total Delay5.4 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
1.6d / 1.0d SLA
Awaiting Parts - Customer
8.0d
Assigned
0.7d / 0.5d SLA
Scheduled
5.6d / 2.0d SLA
Tech On Site
0.9d / 1.0d SLA
Closed
3.1d / 1.0d SLA
L2 Review
5.0d
Invoice Pending
0.1d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
10/20 KWeiss, Reviewed pictures and closing w/o for billing review.
10/2/25 RCook- Spoke with Michael, scheduled for 10/9/25. Arrival time between 8-10AM. Michael did confirm all parts have been received. See logged call on WO-01305523
10/2/25 RCook- LM with Michael Gilbert to schedule and confirm all parts are on site. See logged call on WO-01305523
10/1/25 RCook- SRobinson sent teams message stating ferno has confirmed the customer has received parts for WO-01305523, WO-01305527 & WO-01305528 and they are ready to schedule.
10/1/25 CDempsey- Possibly waiting on parts from Ferno. Should touch base with customer.
9/22/25 SReich - Left a vm on WO-01305523, gave him the WO # and told him someone would be reaching out to schedule. Parts were entered per a teams message from Natosha Beierschmitt.
🔧Work Performed
On the X1, I installed the new patient left caster, provided by Ferno, and made sure it was tight. I then removed the other caster on the right side and replaced the caster with a new one. I then replaced the shoulder bolt for the shock frame and re-installed the red strap. I then replaced the other two wheels. I removed the bent safety bail and installed a new safety bail. I then function tested the cot and all passed. I returned the unit to service.
🔍Technician Findings
On the X1, I arrived and found that the load end patient left wheel, and caster had come off and the patient right caster had worn out bearings. I then discovered that the shock frame bolt had come out and was not holding the red strap on. I found that the other two wheels had flat spots and needed to be replaced. I then found that the safety bail was bent and was not allowing the bail to swing.
⚠️Problem Description
***PART RETURN IS REQUIRED IF RMA IS NOT IN BOX, Email ferno@emsar.com for label***
SN 24C204146
WHEEL CAME OFF. REPLACE WITH NEW WHEEL AND SEND PARTS BACK. RETURN ALL PARTS
Part ID 0806069
Return Parts: Yes
KIT, WHEEL FORK LOAD END 6X2
📄Description
Ferno Power X1 Power Cot
📞
Call
Completed Oct 2, 2025
10/2/25 RCook- Spoke with Michael, scheduled for 10/9/25. Arrival time between 8-10AM. Michael did confirm all parts have been received. See logged call on WO-01305523
📞
Call
Completed Oct 2, 2025
10/2/25 RCook- LM with Michael Gilbert to schedule and confirm all parts are on site. See logged call on WO-01305523
📞
Call
Completed Sep 22, 2025
Left a vm on WO-01305523, gave him the WO # and told him someone would be reaching out to schedule. Parts were entered per a teams message from Natosha Beierschmitt.
Oct 20, 2025 1:12pm
10/20 KWeiss, Reviewed pictures and closing w/o for billing review.
📅
WO-01305527 - Repair - Power X1 - CARROLLTON - ALLEGIANCE MOBILE HEALTH
Oct 9, 2025 1:00pm – 3:00pm
Work Order: WO-01305527
Account: FERNO
Customer Case Number: PCS2501566 Claim #643
Location: ALLEGIANCE MOBILE HEALTH
Address: 1681 S BROADWAY, CARROLLTON, TX 75006
Manufacturer: Ferno
Model #: Power X1
Serial #: 24C204146
Contact Name: Michael Gilbert
Contact Phone: 972-922-5638
Contact Email: mich