Cuyahoga Falls Fire Department - Wyoga Lake Rd. - Inspection Repair
Closed
Q-00030216
Inspection #: 28951
Asset ID: 13243131
Model: Stryker Medical - 6506 PowerPro
Notes: Disinfect and clean cot, mattress, restraints, and accessories.
Inspected cot to confirm the product is free from pinch or cut hazards. All welds and assemblies are intact, hardware and wires are properly secured.
Cycle tested cot in all positions, in both power and manual backup mode.
Loaded and unloaded cot with fastener in both power and manual mode.
Cot fully operational, reviewed inspection results with customer, returned cot to service.
Slider and mag slider worn and rotated, he pl and pr half shell bearings worn and replaced, scheduled psm was completed and then unit was returned to regular service. 34.5hrs Noticeable physical damage on device Parts Needed- Emsar to follow up with quote for needed replacement part. PL XPS wing cracked in half and needs replaced, Tech has part in stock, Part needed quoted is 6500-003-044s X1. I have the part in stock.
Pl xps wing replaced, repair service and evaluation was completed, on site doing pms so no travel involved.
🔍Technician Findings
PL xps wing cracked
34.5hrs
⚠️Problem Description
Q-00030216
Inspection #: 28951
Asset ID: 13243131
Model: Stryker Medical - 6506 PowerPro
Notes: Disinfect and clean cot, mattress, restraints, and accessories.
Inspected cot to confirm the product is free from pinch or cut hazards. All welds and assemblies are intact, hardware and wires are properly secured.
Cycle tested cot in all positions, in both power and manual backup mode.
Loaded and unloaded cot with fastener in both power and manual mode.
Cot fully operational, reviewed inspection results with customer, returned cot to service.
Slider and mag slider worn and rotated, he pl and pr half shell bearings worn and replaced, scheduled psm was completed and then unit was returned to regular service. 34.5hrs Noticeable physical damage on device Parts Needed- Emsar to follow up with quote for needed replacement part. PL XPS wing cracked in half and needs replaced, Tech has part in stock, Part needed quoted is 6500-003-044s X1. I have the part in stock. https://www.ebiotrack.com/workorder.php?wo=68398
📄Description
Stryker 6506 Power-PRO XT Power Cot
Sep 23, 2025 10:56am
@Lawrence Yoder please assign to me im on site doing pms and will take care of this today
📅
WO-01305706 - Repair - 6506 Power-PRO XT - Cuyahoga Falls - Cuyahoga Falls Fire Department - Wyoga Lake Rd.
Sep 23, 2025 3:00pm – 3:15pm
Work Order: WO-01305706
Account: CUYAHOGA FALLS FIRE DEPARTMENT
Customer Case Number:
Location: Cuyahoga Falls Fire Department - Wyoga Lake Rd.
Address: 3497 Wyoga Lake Road, Cuyahoga Falls, OH 44221
Manufacturer: Stryker Medical EMS
Model #: 6506 Power-PRO XT
Serial #: 171040861 truck 216
Contact
Timeline
📞
Case Linked Sep 22, 10:29 AM
Case Number:00529256
Subject:Cuyahoga Falls Fire Department - Wyoga Lake Rd. - Inspection Repair
Status:Closed
📋
▶
2
Work Order Created Sep 23, 9:29 AM
Work Order Created → Status: Ready for Scheduling
📅
Status: Ready for Scheduling by Shelby Robinson
From:Entered
To:Ready for Scheduling
Duration in Previous:0m
👷
▶
2
Technician Assigned Sep 23, 11:02 AM
Technician Assigned → Status: Assigned
👷
Status: Assigned by Lawrence Yoder
From:Ready for Scheduling
To:Assigned
Duration in Previous:1h 32m
🗓️
▶
2
Status: Scheduled Sep 23, 1:26 PM
Status: Scheduled → Dispatch Scheduled
📅
Dispatch Scheduled by Teddie Bracken
Scheduled For:2025-09-23T19:00:00.000+0000
🔩
2x Parts Added Sep 23, 1:34 PM
Type:Parts, Labor
Status:Open
Qty:1, 0.25
🔧
Status: Tech On Site by Teddie BrackenSep 23, 2:51 PM
From:Scheduled
To:Tech On Site
Duration in Previous:1h 24m
✔️
Status: Completed by Teddie BrackenSep 23, 3:17 PM
From:Tech On Site, Completed
To:Completed, Closed
Duration in Previous:25m, 0m
📄
▶
2
Status: Ready to Bill Sep 24, 9:49 AM
Status: Ready to Bill → Status: Invoiced
💰
Status: Invoiced by Diane Patton
From:Invoice Pending
To:Invoiced
Duration in Previous:0m
🚨49.0 business days
🧾
Invoice Created Dec 7, 2:05 PM
Invoice Number:SM-241844
Bottlenecks
Closed
Duration: 1.6 business days (threshold: 1.0 business days)