10-01-HSolano- Travel "To Site" is in WO-01295100.
Please assign to Hermilo Solano
ARP to site is 20 miles. Zone 1
🔧Work Performed
The AIV needed to be replaced due to a microleak that exceeded the test limit for both Pressure and Leak test. The tests passed after replacing the part and purging.
🔍Technician Findings
N/A
⚠️Problem Description
28R
SYS-LC-1200
Repair
Leak and Pressure Test not passing
📄Description
1200 Series Quaternary Pump
📞Call Description
OCT 2025 LC REPAIR
Oct 2, 2025 11:04am
@Hermilo Solano just added your name to this just a heads up
Timeline
📞
▶
4
Case Linked Oct 2, 10:55 AM
Case Linked → Work Order Created → Status: Ready for Scheduling → Technician Assigned
📋
Work Order Created
Work Order:WO-01307438
Type:Repair
Priority:PC
📅
Status: Ready for Scheduling by Leona Coonrod
From:Entered, Ready for Scheduling
To:Ready for Scheduling, Assigned
Duration in Previous:0m
👷
Technician Assigned by Leona Coonrod
Technician:a0hVU000000HkpeYAC, Hermilo Solano
👷
Technician Assigned by Michael BurrisOct 2, 11:01 AM
Technician:Hermilo Solano
⏱️
5x Labor Added Oct 2, 12:12 PM
Type:Labor, Travel, Parts
Status:Open
Qty:2, 0.66666666666667, 25, 1
✔️
▶
2
Status: Completed Oct 2, 12:21 PM
Status: Completed → Status: Closed
🏁
Status: Closed by Hermilo Solano
From:Completed
To:Closed
Duration in Previous:1m
📄
Status: Ready to Bill by Hannah ShawOct 2, 4:02 PM
From:Closed
To:Ready to Bill
Duration in Previous:3h 39m
🚨14.0 business days
📨
▶
2
Status: Invoice Pending Oct 22, 3:27 PM
Status: Invoice Pending → Status: Invoiced
💰
Status: Invoiced by Hannah Shaw
From:Invoice Pending
To:Invoiced
Duration in Previous:2m
⏳4.0 business days
🧾
Invoice Created Oct 27, 12:22 PM
Invoice Number:SM-244382
Bottlenecks
Ready to Bill
Duration: 13.5 business days (threshold: 1.0 business days)