Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems - V100
Notes: Will not power on. Unit\'s battery is inop and needs to be replaced. Reviewed service with customer,
quote needed
1 battery, 2037103-016
2/12 KSmith -- Scheduled for 2/20 with Thalya. 9-11am
2/12 KSmith -- Sent email to schedule for 2/20
2/11/26 RBerry Parts delivered to site on 1/17/26 but WO was not released to ASC until 2/11/26. Updated due date and removed delay code.
🔧Work Performed
Removed and replaced battery from unit. Tested BP and SPo2. Battery must charge 24hrs. note left on device.
🔍Technician Findings
Upon arrival found unit in stated condition.
⚠️Problem Description
Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems - V100
Notes: Will not power on. Unit\'s battery is inop and needs to be replaced. Reviewed service with customer,
quote needed
1 battery, 2037103-016
📧
Email:RE: ELIZABETH CHURCH MANOR -- EMSAR Repair -- WO-01311440
Completed Feb 12, 2026
From: ttauterouff@umhwc.org
To: ksmith@emsar.com
Yes sounds good
From: Kourtney Smith <ksmith@emsar.com>
Sent: Thursday, February 12, 2026 8:29 AM
To: Tauterouff, Thalya <ttauterouff@umhwc.org>
Subject: RE: ELIZABETH CHURCH MANOR -- EMSAR Repair -- WO-01311440
CAUTION: This email originated from outside your organization. Exercise caution when opening attachments or clicking links, especially from unknown senders.
Would Friday, February 20th work for the technician to come
📧
Email:RE: ELIZABETH CHURCH MANOR -- EMSAR Repair -- WO-01311440
Completed Feb 12, 2026
From: ttauterouff@umhwc.org
To: ksmith@emsar.com
Yes I can schedule this. I did email you the other day telling you we have a part here for the Hoyer and also we have acquired 3 lifts from another facility we would like inspected so we can put them out on the floors for use.
Thank you
From: Kourtney Smith <ksmith@emsar.com>
Sent: Thursday, February 12, 2026 8:14 AM
To: Tauterouff, Thalya <ttauterouff@umhwc.org>
Subject: ELIZABETH CHURCH MANOR -- EMSAR Repair -- WO-01311440
📧
Email: ELIZABETH CHURCH MANOR -- EMSAR Repair -- WO-01311440
Completed Feb 12, 2026
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: ELIZABETH CHURCH MANOR -- EMSAR Repair -- WO-01311440
Body:
Hi Thayla,
I’m Kourtney with EMSAR, and I’m reaching out to schedule the onsite services for your healthcare equipment. Would you be able to confirm scheduling for the units at ELIZABETH CHURCH MANOR located at 863 Front St?
· If not, would you have the contact information of the person I need to speak with?
· If you are the correc
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Feb 9, 2026
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Feb 2, 2026
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Jan 26, 2026
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Jan 19, 2026
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Jan 12, 2026
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Jan 5, 2026
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Dec 29, 2025
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Dec 22, 2025
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Dec 15, 2025
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Dec 8, 2025
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Dec 1, 2025
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Nov 24, 2025
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Nov 17, 2025
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Nov 10, 2025
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Nov 3, 2025
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Oct 27, 2025
To: ttauterouff@umhwc.org
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01311440Location: ELIZABETH CHURCH MANORAddress: 863 FRONT ST BINGHAMTON, NY 13905Device: Problem Description: Q-00030717
Inspection #: 10141
Asset ID: 13037386
Model: Ge Medical Systems
📅
WO-01311440 - Repair - BINGHAMTON - ELIZABETH CHURCH MANOR
Feb 20, 2026 9:30am – 12:00pm
Work Order: WO-01311440
Account: ELIZABETH CHURCH MANOR
Customer Case Number:
Location: ELIZABETH CHURCH MANOR
Address: 863 FRONT ST, BINGHAMTON, NY 13905
Manufacturer:
Model #:
Serial #:
Contact Name: Thalya Tauterouff
Contact Phone: 607-722-3463 Ext. 3230
Contact Email: ttauterouff@umhwc.org
Timeline
📞
Case Linked Oct 21, 12:55 PM
Case Number:00537157
Subject:ELIZABETH CHURCH MANOR - Inspection Repair