11/21 Up to today SO can't be added in SM. WO will be closed without part
<D
11/08 An email was sent to add parts in SM. MD
11/06 Google map was used to calculate zone 2. MD
Please assign to Maria Davis ARP to site is 65 miles. Zone 2
🔧Work Performed
11/10/2025 After following up customer stated "Yes, just did it this morning and it looks great. All good".
11/07/2025 After following up customer stated "QC doesn't look good, there is no drug towards the end" Working on the phone with Jesika, to check the column in the MSD side and that likely the length of the column with extraction source and white tip is know for this behavior in toxicologic labs. Explained here step by step how to re-do it. She will try on 11/10
11/06/2025 GCMS was tuned and vented. Ion source and dry pump were cleaned. GC parts were replaced. Per costumer request Column was replaced, system was pumped down and stabilized for sometime until H2O and N2 were below the limit. Customer's method was loaded and stabilized and customer ran a negative sample and looked OK, still customer will run a sequence to test QC samples.
🔍Technician Findings
11/21 Up to today SO can't be added in SM. WO will be closed without part
⚠️Problem Description
28R
SYS-GM-5977T
Contract ends 07/31/2026 repair included
PM
3540A
US2107A005 Agilent 8890 GC System Custom
G7077-80057
MY2104S014 New-24V IDP3 scroll pump with valve
G7077B
US2109M040 5977B Inert Plus MSD Turbo EI Mainframe
G4514A
RO21037050 7693A Tray, 150 Vial
G4513A
RO21045060 7693A Autoinjector
G3397B
JBDR Ion Gauge Controller kit for 5977
📄Description
Agilent - 8890 GC System Custom
Nov 19, 2025 1:38pm
@Maria Davis Per your notes, system is up and running. Why is this not closed?