Troubleshoot Noise:
Clinic is reporting they have an exam table that is screeching when it goes
up and down.
Address is: 323 Soquel Way, Sunnyvale, CA 94085
Room:Tempest
Serial number: V178674
625 Midmark Exam table
After troubleshooting, I found that the outer shroud wasn’t installed correctly. The end lip is supposed to slide under the lip of the shroud beneath it. I reinstalled the shroud properly and tested the chair several times in the up and down positions, confirming that the screeching noise is no longer present.
🔍Technician Findings
Customer led me to the room with the exam chair making a screeching noise when moving up or down. Confirmed the correct unit by serial number and verified the noise during testing.
⚠️Problem Description
Troubleshoot Noise:
Clinic is reporting they have an exam table that is screeching when it goes
up and down.
Address is: 323 Soquel Way, Sunnyvale, CA 94085
Room:Tempest
Serial number: V1786734
625 Midmark Exam table https://www.ebiotrack.com/workorder.php?wo=69467
📋
Outbound to +19252346755
Completed Nov 5, 2025