Midmark Service Request Dispatch Authorization - WO-00126191
Closed
Travel Zone 1
Dispatch Type Repair
PROBLEM SUMMARY
NWD C981, C232, C571, C103, C981 Element heating up on its own.
ADDITIONAL NOTES
If part(s) are identified after speaking with Midmark Tech Service, the Service Company Tech must contact the POC for
authorization before ordering. Once approved, notify Midmark immediately. If more time is needed for the customer’s decision
to proceed, inform Midmark Dispatch at 937.688.9175 or dispatch@midmark.com. For assistance, troubleshooting, or parts
identification, contact Veterinary Tech Service at 844-856-1232, option 1.
PCR CODES
Problem Cause Resolution
HIGH TEMPERATURE C54X C57X C98X BOARD FAULTY REPLACE PARTS
In Transit Notification, Tracking Number 1ZE444250346631445
Estimated deliveryToday, December 08 between 4:30 P.M. - 8:30 P.M.
12/3/25 NSuarez SRA attached is not filled out please fill it out.
12/2/25 SReich
Order No: 740733
Customer PO No: SR1488800
Customer: Tanasbourne Veterinary Emergency
Service Request NO: 1488800
Ship To Name: EMSAR Inc
Ship To Address: 4877 13th Ave NE
Ship To City/State: Keizer, OR 97303-3813
Waybill (Tracking #): 1ZE444250346631445
Carrier: UPS-Parcel-Ground
The following items have shipped from Midmark Corporation on: 01-DEC-2025
LINE NO: 1.1 QTY: 1 ITEM: 002-10070-00 SERIAL #: DESCRIPTION: PM KIT FOR M11
11-18 Mburris called site set appt with Rose
11-18 Mburris
1ZE444250372967301
Estimated delivery
Thursday, November 20 between 10:00 A.M. - 1:00 P.M
11/13/2025jsells
Ship To Name: Tanasbourne Veterinary Emergency
Ship To Address: 17400 NW Corridor Ct
Ship To City/State: Beaverton, OR 97006-3557
Waybill (Tracking #): 1ZE444250372967301
Carrier: UPS-Parcel-Ground
The following items have shipped from Midmark Corporation on: 13-NOV-2025
LINE NO: 1.1 QTY: 1 ITEM: 002-2063-00 SERIAL #: DESCRIPTION: KIT, 115V M9/M11 CONTROL PCB
11/12/2025 JRolston - Called site POC, Rachel Rose @ (503) 629-5800. Advised her that dispatch has been received, and that Coordinators will call to schedule accordingly once tracking / ETA for parts has been determined.
🔧Work Performed
Replaced main board. Installed the PM kit. Ran a test cycle to verify working.
🔍Technician Findings
C981, C232, C571, C103 errors - board causing elements to overheat, dam has melted and stuck to the drum.
⚠️Problem Description
Travel Zone 1
Dispatch Type Repair
PROBLEM SUMMARY
NWD C981, C232, C571, C103, C981 Element heating up on its own.
ADDITIONAL NOTES
If part(s) are identified after speaking with Midmark Tech Service, the Service Company Tech must contact the POC for authorization before ordering. Once approved, notify Midmark immediately. If more time is needed for the customer’s decision to proceed, inform Midmark Dispatch at 937.688.9175 or dispatch@midmark.com. For assistance, troubleshooting, or parts identification, contact Veterinary Tech Service at 844-856-1232, option 1.
PCR CODES
Problem Cause Resolution
HIGH TEMPERATURE C54X C57X C98X BOARD FAULTY REPLACE PARTS
📄Description
Midmark Model M11 (Series) Self Contained Steam Sterilizer (Includes ALL M11 Models)
📞
Call
Completed Nov 25, 2025
Jason reached out via Teams. am on a non warranty Midmark M11 call. I used a PM kit from my trunk stock but midmark charged the customer and is sending a new one to my house. DO I still mark down that I used one or will that cause accounting issues?
Recommended adding part as customer supplied and to ensure replacement part is received.
📋