Note
11/17/2025 12:37:24 MOHPANDE
[11-17-2025 ATI RM TEAM Mohit Pandey]
This is related to part VSFLDDP qty 1 ordered on SO 6007927419 under STO 4104782978, as the part is 0 stock in Memphis and will have to be backordered.
Kindly confirm, how soon do you need the part because part needs to be hotline/procure from Germany warehouse. If it is not urgent, we will follow the process and expedite it to get it sooner from elsewhere.
Hello Eng-Siang,
Please advise the best ETA to the LC for normal supplied material.
____________________
Note
11/17/2025 09:10:47 WIGEORGE
field pm
____________________
Note
10/29/2025 09:49:53 CHIR
Customer issued PO for work to be delivered in 2026. Setting to Deliver Solution with due by December 2026. We expect customer to contact us when they want the PM.
____________________
Note
07/08/2025 11:03:58 WF-BATCH
Agent: WIGEORGE Timestamp: 2025-07-08 11:03:47
trade services; not a real quotation; this is an estimate to be used for their 2026 budget
07/14/2025 Shubham Srivastava
QOF/HCPO/FSP PO Received on : 07/14/2025
PO# 4100068303
NTE Amount: USD 3,195.05
Response/Terms: STD / Net 90 Days
Tax Status: Taxable
Invoicing Address :
St Jude Medical Cardiology Division Inc
100 Abbott Park Rd
ABBOTT PARK IL 60064-3502
USA
Please see attachment for details/ PO attached at SR/SO level—Routed to WLA. PO Acknowledgement sent.
12.18.25 hshaw - removed from SM-249149 and will rebill.
12.18.25 Hshaw - zone 6 is missing
12-2 Mburris emailed back to POC Dan that that date and time is still good
12/2/25 lcoonrod
Per Daniel: Yea we got the parts. Can we get the tech scheduled on the calendar? We had 12/11 tentatively penciled in, does that still work by any chance?
12/1/25 lcoonrod sent email to confirm receipt of parts.
11/18/25 lcoonrod part VSFLDDP has not shipped yet.
🔧Work Performed
Completed Service Tasks:
Inspection and Cleaning
Performed a comprehensive inspection and cleaning of:
Spectrometer
Valve block
Manifold
Component Replacement:
Replaced ion source and filaments
Installed new O-rings and seals where required
System Verification:
Verified integrity of the vacuum system
Performed helium sensitivity check
Completed full system calibration
Calibrated Leak Maintenance:
Replaced the calibrated leak
Supplied ISO traceable certificate for the new calibrated leak
🔍Technician Findings
As found: Unit in work lab
⚠️Problem Description
Note
11/17/2025 12:37:24 MOHPANDE
[11-17-2025 ATI RM TEAM Mohit Pandey]
This is related to part VSFLDDP qty 1 ordered on SO 6007927419 under STO 4104782978, as the part is 0 stock in Memphis and will have to be backordered.
Kindly confirm, how soon do you need the part because part needs to be hotline/procure from Germany warehouse. If it is not urgent, we will follow the process and expedite it to get it sooner from elsewhere.
Hello Eng-Siang,
Please advise the best ETA to the LC for normal supplied material.
____________________
Note
11/17/2025 09:10:47 WIGEORGE
field pm
____________________
Note
10/29/2025 09:49:53 CHIR
Customer issued PO for work to be delivered in 2026. Setting to Deliver Solution with due by December 2026. We expect customer to contact us when they want the PM.
____________________
Note
07/08/2025 11:03:58 WF-BATCH
Agent: WIGEORGE Timestamp: 2025-07-08 11:03:47
trade services; not a real quotation; this is an estimate to be used for their 2026 budget
07/14/2025 Shubham Srivastava
QOF/HCPO/FSP PO Received on : 07/14/2025
PO# 4100068303
NTE Amount: USD 3,195.05
Response/Terms: STD / Net 90 Days
Tax Status: Taxable
Invoicing Address :
St Jude Medical Cardiology Division Inc
100 Abbott Park Rd
ABBOTT PARK IL 60064-3502
USA
Please see attachment for details/ PO attached at SR/SO level—Routed to WLA. PO Acknowledgement sent.
📄Description
VS PD03 Portable Leak Detector
📞Call Description
VPD Field PMTrade
📧
Email: WO-01315911 6035 Stoneridge Dr Pleasanton, CA 94588
Completed Dec 2, 2025
To: daniel.watnick@abbott.com
CC:
BCC: michael.burris@emsar.com
Attachment: --none--
Subject: WO-01315911 6035 Stoneridge Dr Pleasanton, CA 94588
Body:
Hi,
Dan yes, we are still good for 12-11 between 7am and 8:30am
Best Regards,
Michael Burris Jr
Account Service Coordinator I
Work Phone : (937) 802-9367
michael.burris@emsar.com
www.emsar.com
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Dec 2, 2025
To: daniel.watnick@abbott.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01315911Location: Abbott LaboratoriesAddress: 6035 Stoneridge Dr Pleasanton, CA 94588Device: MY14120001Problem Description: Note
11/17/2025 12:37:24 MOHPANDE
[11-17-2025 ATI RM TEAM
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Nov 25, 2025
To: daniel.watnick@abbott.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01315911Location: Abbott LaboratoriesAddress: 6035 Stoneridge Dr Pleasanton, CA 94588Device: MY14120001Problem Description: Note
11/17/2025 12:37:24 MOHPANDE
[11-17-2025 ATI RM TEAM
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Nov 18, 2025
To: daniel.watnick@abbott.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01315911Location: Abbott LaboratoriesAddress: 6035 Stoneridge Dr Pleasanton, CA 94588Device: MY14120001Problem Description: Note
11/17/2025 12:37:24 MOHPANDE
[11-17-2025 ATI RM TEAM
Dec 15, 2025 11:15am
@MICHAEL MCNAMARA I have reviewed and moved WO to RTB.
Dec 18, 2025 5:37pm
@Kwame Cobb please enter your mileage on 12/10. Thank you @David Carvalho