Cot fully operational, reviewed inspection results with customer, returned cot to service.
cx needs quote for leg bearings 16 (6500-001-327) Follow-up required. Parts Needed- Emsar to follow up with quote for needed replacement part. Noticeable physical damage on device
12/10 KKelch-FSE reached out to Tim and scheduled directly with the site
12/4/2025 Sramen - Adjusted due date,parts shipping on 1315519. Need to schedule together.
11/26/25 RBerry- Received confirmation from DCarvalho to send Chris Womack. Assigned WO and sent teams chat to KKelch.
11/26/25 RBerry- Found note in Special instructions that site is requesting Chris Womack from TX. Returned to dispatch to clear tech name until confirmed with FOM.
🔧Work Performed
On the 6506, I removed the legs and replaced the 16 half shells. I then removed the sensor housings and replaced the crest-to-crest springs, slider, and slider magnets. I then replaced the Left and right slider housings. I then removed the leaking hose and installed a new 3/8 hose. I then installed a new backrest piston and safety bail. I installed three new zip ties. I function-tested the unit and all passed. I returned the cot to the service. NO TRAVEL CHARGED.
🔍Technician Findings
On the 6506, I arrived and found that the cot was struggling to raise. The bearings were worn. The slider and slider magnet were damaged, and the crest-to-crest springs were broken. The left and right Slider housings were grooved and needing to be replaced. I then discovered that the large hose was leaking, the safety bail was worn, and the backrest piston was spongy. The zip ties for the steer cable were missing.
⚠️Problem Description
Cot fully operational, reviewed inspection results with customer, returned cot to service.
cx needs quote for leg bearings 16 (6500-001-327) Follow-up required. Parts Needed- Emsar to follow up with quote for needed replacement part. Noticeable physical damage on device https://www.ebiotrack.com/workorder.php?wo=69865
📄Description
Stryker 6506 Power-PRO XT Power Cot
📧
Email: Update: Awaiting Parts for Your Work Order
Completed Nov 21, 2025
To: tim@midamericaambulancerepair.com
CC:
BCC:
Attachment: --none--
Subject: Update: Awaiting Parts for Your Work Order
Body:
Dear Customer,Thank you for choosing EMSAR for your service needs.We are currently awaiting the parts necessary to complete the work for your device.Work Order #: WO-01316582Location: Southern Paramedic Service EMS - Brinkley ARAddress: 1100 N Main St. Brinkley, AR 72021Device: 6506 Power-PRO XT 161141405Problem Description: Cot fully oper
📅
WO-01316582 - Repair - 6506 Power-PRO XT - Brinkley - Southern Paramedic Service EMS - Brinkley AR
Dec 16, 2025 10:00am – 4:00pm
Work Order: WO-01316582
Account: SOUTHERN PARAMEDIC SERVICE EMS
Customer Case Number:
Location: Southern Paramedic Service EMS - Brinkley AR
Address: 1100 N Main St., Brinkley, AR 72021
Manufacturer: Stryker Medical EMS
Model #: 6506 Power-PRO XT
Serial #: 161141405
Contact Name: Tim Kalvin
Contact
Timeline
📞
Case Linked Nov 19, 11:32 PM
Case Number:00545448
Subject:L2 review
Status:Closed
📋
▶
2
Work Order Created Nov 20, 3:19 PM
Work Order Created → Status: L2 Screening
🔍
Status: L2 Screening by Jessica Sells
From:Entered
To:L2 Screening
Duration in Previous:0m
📦
▶
2
Parts Requested Nov 20, 3:56 PM
Parts Requested → Status: Awaiting Parts
📦
Status: Awaiting Parts by Keith Weiss
From:L2 Screening
To:Awaiting Parts
Duration in Previous:35m
⏳4.0 business days
📅
Status: Ready for Scheduling by Sean WeaverNov 25, 11:01 AM