Work Order 360 View
Detailed work order timeline, KPIs, and analysis
PC
SLA Clock
Per client agreement · Target: Dec 3, 2025
✓ SLA Met (4 biz days early)
🕐 Clock stopped: Completed (Nov 25, 2025)
▶ Clock Running
Status: Created
Nov 25 → Nov 25, 2025
0 business days (counted)
Running total: 0 of ? biz days used
▶ Clock Running
Status: Ready for Scheduling
Nov 25 → Nov 25, 2025
0 business days (counted)
Running total: 0 of ? biz days used
0d
▶ Clock Running
Status: Assigned
Nov 25 → Nov 25, 2025
0 business days (counted)
Running total: 0 of ? biz days used
0d
▶ Clock Running
Status: Tech On Site
Nov 25 → Nov 25, 2025
0 business days (counted)
Running total: 0 of ? biz days used
Created: Nov 25, 2025
Completed: Nov 25, 2025
Show SLA Clock Detail (4 periods)
Period
Status
Clock
Biz Days
Running Total
Nov 25 → Nov 25
Created
▶ Running
0
0
Nov 25 → Nov 25
Ready for Scheduling
▶ Running
0
0
Nov 25 → Nov 25
Assigned
▶ Running
0
0
Nov 25 → Nov 25
Tech On Site
▶ Running
0
0
Coordinators
0.6 days
Over SLA
Billing
6.7 days
Over SLA
Created: Nov 25, 2025
Due: Dec 3, 2025
Completed: Nov 26, 2025 (7 days early)
⏱️
6.7d in Ready to Bill
Longest Stage
🔄
7 transitions
Status Changes
📦
0 orders, 0 quotes
Parts & Quotes
👤
Self-Scheduled
Scheduling
AI Analysis
Powered by OpenAI
Generating AI analysis...
🎫
Case #00546693
6008242748
Closed
15K
SYS-LC-1260II
Repair
WO81912378:BINARY PUMP LEAKING
Contact
Diana Hernandez Guerrero
🚨
Primary Delay
6.7 days stuck in "Ready to Bill"
⏱️
Total Delay
5.8 business days over SLA thresholds
SLA vs Actual Time by Status
Tech On Site
0.2d
/ 1.0d SLA
Completed
0.2d
/ 1.0d SLA
Ready to Bill
6.7d
/ 1.0d SLA
Invoice Pending
0.0d
/ 2.0d SLA
💬 Activities, Notes & Communications
📝 WO Notes (6)
🔒
Internal CSA Notes
Please assign to Jared Rappaport
ARP to site is 17 miles. Zone 1
🔧
Work Performed
Repaired binary pump (Agilent).
🔍
Technician Findings
Following pump maintenance under SO 6008029829, a leak was observed at the binary pump. Identified the mixing capillary and the mixing-capillary–to–pressure-sensor capillary as defective. Replaced both capillaries with like-for-like parts. Performed pump leak-rate test and system pressure test; both passed Agilent specifications at max pressure. Refer to the PM SO for detailed test results.
⚠️
Problem Description
15K
SYS-LC-1260II
Repair
WO81912378:BINARY PUMP LEAKING
📄
Description
1260 Infinity II Binary Pump
📞
Call Description
DEC 2025 LC REPAIR
Timeline
📞
📋
Work Order:
WO-01317161
Type:
Repair
Priority:
PC
📅
From:
Entered, Ready for Scheduling
To:
Ready for Scheduling, Assigned
Duration in Previous:
0m
👷
Technician:
a0hVU000000hjebYAA, Jared Rappaport
🔧
From:
Assigned
To:
Tech On Site
Duration in Previous:
4h 39m
🔩
Type:
Parts
Status:
Open
Qty:
1
⏱️
✔️
From:
Tech On Site
To:
Completed
Duration in Previous:
1h 27m
⏱️
🏁
From:
Completed
To:
Closed
Duration in Previous:
1h 22m
📄
From:
Closed
To:
Ready to Bill
Duration in Previous:
17h 42m
📨
💰
From:
Invoice Pending
To:
Invoiced
Duration in Previous:
3m
🧾
Invoice Number:
SM-248486
Bottlenecks
Assigned
Duration: 4.7 business hours (threshold: 4.0 business hours)
Exceeded by 0.7 business hours
Ready to Bill
Duration: 6.7 business days (threshold: 1.0 business days)
Exceeded by 5.7 business days
ℹ️ Work Order Details
Technician
Jared Rappaport
Work Details (5)
Name
Type
Status
Qty
Amount
Date
WL-03020128
Parts
Open
1
$0.00
Nov 25, 5:48 PM
WL-03020129
Parts
Open
1
$0.00
Nov 25, 5:48 PM
WL-03020149
Labor
Open
1.7
$221.00
Nov 25, 6:01 PM
WL-03020231
Labor
Open
0.88333333333333
$0.00
Nov 25, 7:25 PM
WL-03020232
Travel
Open
17
$0.00
Nov 25, 7:25 PM