Requested work order from dispatch on Dec 5th. Sent 1st request to David on Dec 3rd
🔧Work Performed
Upon arrival i met Keith in ER. Found Prism lift model 625 (SN C6P006461) with dead batteries due to it not being on charging station for unknown period of time. Put in docking station and left to charge. I was asked to look at a Guldmann GX2 lift with a gap in rail with permission from Clint. Drove to other building and found 3/8in gap in rail union. Found one of he bracket bolts loose. Loosened other side and closed gap, then securing and tightening bolts. (GH2 S/N 15822011). Went back to ER room 1E145 and unit batteries were at 20%. I tested all functions of corded remote and all worked. No noise noted coming from unit. I positioned motor by moving without drive function and heard various noises coming from rail. These are due from motor being repositioned with out using drive motor function since drive wheels will not roll when doing this. Unit is working properly but still charging when I left.
🔍Technician Findings
Lift dead upon arrival
⚠️Problem Description
Noisey when moving lift, Hand control not working
PDF
R110123041 (002) complete
Dec 8, 2025
Timeline
📞
Case Linked Dec 4, 7:22 AM
Case Number:00548717