EMS L2 Review * No quote until parts are determined
Closed
Truck 57 POWER LOAD SYSTEM
Stryker Medical - 6390 Power-LOAD
Serial #: 150240281 - Asset ID: 13212213
Erwin was out here last week and did a PM on the powerload , with no issues at that time and it is running into the issue with the bar not going in with the trolley. I told him last week that when they did have the issue we had to pull on the 2 red triggers on the bar and that would make it work. Erwin had no issue last week and now it is doing it again.
WE NEED THIS TRUCK INSERVICE. Can Erwin come to Niagara Active base 6010 Lockport Rd Niagara Falls NY early next week?
The technician will need to evaluate the unit and determine if parts are needed for repairs.
12/15 KSmith -- FSE scheduled for Thursday with Donna
🔧Work Performed
Removed trolley and transfer assemblies. Cleaned and inspected transfer lock plates. Plates show signs of wear and will recommend replacement. Transfer lock trigger slides inspected. Transfer lock pin assembly moves freely. Rollers on anchor show signs of heavy use. (pics attached). Inspection of trolley found the 2 screws that attach the trolley stop ramp were loose (1 totally unscrewed from ramp). Reattached trolley ramp stop and secured with screws. Reassembled transfer/trolley and tested. Transfer now slides in normally. Demonstrated to staff.
🔍Technician Findings
Confirmed transfer does not retract when trolley is pushed in. Have to use manual slides to move transfer.
⚠️Problem Description
Truck 57 POWER LOAD SYSTEM
Stryker Medical - 6390 Power-LOAD
Serial #: 150240281 - Asset ID: 13212213
Erwin was out here last week and did a PM on the powerload , with no issues at that time and it is running into the issue with the bar not going in with the trolley. I told him last week that when they did have the issue we had to pull on the 2 red triggers on the bar and that would make it work. Erwin had no issue last week and now it is doing it again.
WE NEED THIS TRUCK INSERVICE. Can Erwin come to Niagara Active base 6010 Lockport Rd Niagara Falls NY early next week?
The technician will need to evaluate the unit and determine if parts are needed for repairs. https://www.ebiotrack.com/workorder.php?wo=70355
📄Description
Stryker 6390 Power-LOAD Fastener System
📅
WO-01326500 - Repair - 6390 Power-LOAD - Niagara Falls - Mercy Flight WNY - Niagara Active Hose Company
Dec 18, 2025 12:00pm – 2:00pm
Work Order: WO-01326500
Account: MERCY FLIGHT WNY
Customer Case Number:
Location: Mercy Flight WNY - Niagara Active Hose Company
Address: 6010 Lockport Rd., Niagara Falls, NY 14305
Manufacturer: Stryker Medical EMS
Model #: 6390 Power-LOAD
Serial #: 150240281
Contact Name: Donna Cox
Contact Phone:
JPG
trolley 3
Dec 18, 2025
JPG
trolley 4
Dec 18, 2025
JPG
transfer rollers 3
Dec 18, 2025
JPG
trolley 2
Dec 18, 2025
JPG
transfer rollers 4
Dec 18, 2025
JPG
transfer rollers 2
Dec 18, 2025
JPG
trolley 1
Dec 18, 2025
JPG
transfer rollers 1
Dec 18, 2025
Timeline
📞
Case Linked Dec 12, 11:29 AM
Case Number:00551331
Subject:EMS L2 Review * No quote until parts are determined