Midmark Service Request Dispatch Authorization - WO-00138496
Closed
Zone 1
WD-no base function
ADDITIONAL NOTES
For assistance, troubleshooting, or parts identification, contact Medical Tech Service at 844-856-1230. Thank you!
Ship Exchanged Open 744750 015-3481-00 1.00000
Ship To Name: Shoreline OBGYN
Ship To Address: 4 Shaws Cv Ste 204
Ship To City/State: New London, CT 06320-4956
Waybill (Tracking #): 1ZE444250370396708
Carrier: UPS-Parcel-Ground
The following items have shipped from Midmark Corporation on: 26-DEC-2025
LINE NO: 1.1 QTY: 1 ITEM: 002-10050-00 SERIAL #: DESCRIPTION: KIT, FOOT CONTROL ASS'Y
12/18 KSmith -- Scheduled for 12/23 with Jennifer. 8-10am
12/17/2025jsells
Ship To Name: Shoreline OBGYN
Ship To Address: 4 Shaws Cv Ste 204
Ship To City/State: New London, CT 06320-4956
Waybill (Tracking #): 1ZE444250347268195
Carrier: UPS-Parcel-Ground
The following items have shipped from Midmark Corporation on: 17-DEC-2025
LINE NO: 1.1 QTY: 1 ITEM: 015-3481-00 SERIAL #: DESCRIPTION: I/O COMMUNICATION HARNESS
12/17/25 SReich - Left a vm, gave her the WO # and told her someone would be reaching out to schedule around the arrival of parts.
🔧Work Performed
**R&R IO comms harness. After install base would still not function. I tested all the fusesto cinfirm none were blown. All good. Reseated all connections to confirm no loose connections. Spoke with Jason @MM had me plug in another controller and unit worked and operated as it should. New controller was ordered and shipped directly to client.
🔍Technician Findings
Unit OOS
⚠️Problem Description
Zone 1
WD-no base function
ADDITIONAL NOTES
For assistance, troubleshooting, or parts identification, contact Medical Tech Service at 844-856-1230. Thank you!
Ship Exchanged Open 744750 015-3481-00 1.00000