Power is intermittent for Footrest.
Technician to complete product review,eval,and testing to determine if parts are needed
1/12/2026 CRamen spoke with Walt and he requested Friday 1/16/2026 with tech arrival between 9-11am. Unit is located in the OPP3 Building (Specialty Services Pavillion) in Room 2227. Walt is contact 216-870-5549
🔧Work Performed
checked voltage at the transformer going in and going out. Both were at a acceptable range. Confirmed actuator was working by switching tilt and foot rest harness. Ohmed out the motor for the foot rest and was within a acceptable range. All sensors checked and are working properly. Recalibrated the chair but unit fails part way through each time. Issues seems to point towards a board issue. The board is no longer manufactured and are unable to get for the chair. Unit can not be repaired
🔍Technician Findings
power intermittent to footrest
⚠️Problem Description
Q-00031837
Midmark 630-009
V60422
Power is intermittent for Footrest.
Technician to complete product review,eval,and testing to determine if parts are needed
📄Description
Midmark Model: 630 (Series) Human Form Procedures Chair
📋
Inbound from +12168705549
Completed Jan 20, 2026
wanting an inspection report
📋
Inbound from +12168705549
Completed Jan 16, 2026
Transferred call to the coordinator to update customer with tech ETA.
📋
Outbound to +12168705549
Completed Jan 12, 2026
1/12/2026 CRamen spoke with Walk and he requested Friday 1/16/2026 with tech arrival between 9-11am. Unit is located in the OPP3 Building (Specialty Services Pavillion) in Room 2227. Walt is contact 216-870-5549
📋
Outbound to +14136275735
Completed Jan 2, 2026
Called customer back to get more info. He provided issue and asset ID.
📋
Inbound from +12168705549
Completed Jan 2, 2026
Needs quote for Midmark bed
📅
WO-01337351 - Evaluation - 630 (Series) - CLEVELAND, - MetroHealth Medical Center
Jan 16, 2026 9:00am – 10:00am
Work Order: WO-01337351
Account: MetroHealth Medical Center
Customer Case Number:
Location: MetroHealth Medical Center
Address: 2500 METROHEALTH DR Rm 2205, CLEVELAND,, OH 44109
Manufacturer: Midmark Corp
Model #: 630 (Series)
Serial #: V60422
Contact Name: Walt Stepanek
Contact Phone: (21