PROBLEM SUMMARY
WD-door is jammed and not opening
ADDITIONAL NOTES
For assistance, troubleshooting, or parts identification, contact Medical Tech Service at 844-856-1230. Thank you!
Primary Delay2.9 days stuck in "Ready for Scheduling"
⏱️
Total Delay3.0 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
2.9d / 1.0d SLA
Assigned
0.0d / 0.5d SLA
Scheduled
0.6d / 2.0d SLA
Tech On Site
0.2d / 1.0d SLA
Completed
1.0d / 1.0d SLA
Closed
2.0d / 1.0d SLA
Ready to Bill
1.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
2-18 Mburris called and set with Beth 2-18 11am
2/13/2026jsells
Ship To Name: Praxis Health
Ship To Address: 9555 SW Barnes Rd Ste 201
Ship To City/State: Portland, OR 97225-6654
Waybill (Tracking #): 1ZE444250171345363
Carrier: UPS-Air-Next Day Air
The following items have shipped from Midmark Corporation on: 13-FEB-2026
LINE NO: 1.1 QTY: 1 ITEM: 002-10896-00 SERIAL #: DESCRIPTION: KIT, M9/M11-05X DOOR MOTOR ASSY
2/13/26 lcoonrod
only has options for providers or patients. cannot reach anyone. will send email.
🔧Work Performed
Ran motor cycle 10 times with no issues. Reseated M9 cassette. Filled reservoir with water and performed full cycle with no issues.
🔍Technician Findings
Cycle ran without water causing door to lock. M9 tray cassette out of position causing binding on door/seal pressure.
⚠️Problem Description
Travel Zone 1
Dispatch Type Repair Email mzjeylani@gopxh.com
PROBLEM SUMMARY
WD-door is jammed and not opening