FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R4.5 and communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully
functional”.
🔍Technician Findings
Template Name:
Template Description:
Problem Found:
Action Taken:
“Completed FMI 78101 by installing SW Version R4.5 and communicated with customer to scrap previous software versions”.
Verification Test:
“Performed Probe and Connectors Checks for all probes. Confirmed all probes pass. Completed all Peripheral Checks-system passed.
Completed full system diagnostics-system passed. Equipment passed all required tests and has been returned to the customer fully
functional”.
⚠️Problem Description
FMI 78101D: Venue Go Thermal Event due to battery Purpose: Each affected system will have a SW Version R4.5 update installed. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at TCS.FMIAdmin@gehealthcare.com or 8668836784 ? OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare - Venue Go Ultrasound
📞Call Description
FMI 78101
📅
WO-01352161 - FCO - Venue Go Ultrasound - NISKAYUNA - GE GLOBAL RESEARCH
Feb 17, 2026 12:00pm – 1:00pm
Work Order: WO-01352161
Account: GE Healthcare FMI
Customer Case Number: 1-554880306179, 1-554908016481
Location: GE GLOBAL RESEARCH
Address: 1 RESEARCH CIR, NISKAYUNA, NY 12309
Manufacturer: GE Healthcare
Model #: Venue Go Ultrasound
Serial #: VGD002718
Contact Name:
Contact Phone:
Contact Email: