08/14/2025 ALuman - dispatch to Raymond Ebanks to complete. Email sent to Justin for scheduling confirmation for 8/22
08/11/2025 DCox - Quote has been accepted. Planner to schedule for FSE to complete
08/05/2025 EMartin PWO pending entitlement FSE to work with POC to get hard copy PO and quote updated
🔧Work Performed
Travel to site. Replaced the DA PCBA. Device still failed system leak test. Removed new DA PCBA & replaced with the old. Device passed all other tests except system leak. Pressure test tubing & inspected rubber kits for holes or cracks. Returned DA PCBA. Ordered Blower assembly, Tubing & rubber kits.
🔍Technician Findings
Repair
⚠️Problem Description
Device failed System Leak Test during annual PM.
📄Description
V60 Ventilator
📞Call Description
Repair - Trade
Aug 5, 2025 9:05am
Hi, @Ashley Luman. This is the trade repair requested per the below. Thanks!
WO-13274383
Delray Medical Center
5352 Linton Blvd
Delray Beach, FL 33484
📅
WO-01295776 - Repair - V60 Vent - Delray Beach - Delray Medical Center
Sep 4, 2025 8:00pm – 8:00pm
Work Order: WO-01295776
Account: PHILIPS HEALTHCARE
Customer Case Number: WO-13274383
Location: Delray Medical Center
Address: 5352 Linton Blvd, Delray Beach, FL 33484
Manufacturer: Philips Healthcare
Model #: V60 Vent
Serial #: 100196130
Contact Name: Justin Lazore
Contact Phone: 561-495-3227
Conta
Case #00516309 – URGENT Philips WO Request
Status: Closed | Priority: Standard
💬Case Comment
Internal Aug 5, 2025 9:05am
08/05/2025 JRolston - Created below EWO's accordingly.
WO-01295773
WO-01295776