Q-00030269
**Darrell Hendrix has the parts needed for this service.
This customer has three cot that are in need of a repair. Need work orders generated (see pictures in email)
SN 161039457 the cot is not wanting to lift up
Possible parts needed for repair
16 half shell bearings,
6500-101-327
1 qt. fluid, Mobil 1
SN 170139211
The emergency release handle on unit is broken
Possible parts needed
1 lever, 6500-001-141
1 spring, 6500-001-147
1 pin, 6500-001-146
1 actuator, 6500-001-140
1 screw, 0023-163-000
SN 170139210
The XPS side rail is broken here are parts I feel are needed to fix the cot
6500-003-084 x2
0004-402-000 x4
6500-003-102 x2
0004-400-000 x4
6500-003-120 x2
6500-003-045 x2
6500-003-085 x2
I found said unit. I found was struggling going up I found parts that was broken or worn. I then used part # 6500-101-327 X16 half shell. I then cleaned and disinfected the unit and returned to service.
🔍Technician Findings
Found that the cot is struggling going up
⚠️Problem Description
Q-00030269
**Darrell Hendrix has the parts needed for this service.
1 of 3
This customer has three cot that are in need of a repair. Need work orders generated (see pictures in email)
SN 161039457 the cot is not wanting to lift up
Possible parts needed for repair
16 half shell bearings,
6500-101-327
1 qt. fluid, Mobil 1
Work Order: WO-01306209
Account: OKMULGEE COUNTY EMS
Customer Case Number:
Location: OKMULGEE COUNTY EMS
Address: 1213 E. 20th Street, Okmulgee, OK 74447
Manufacturer: Stryker Medical EMS
Model #: 6506 Power-PRO XT
Serial #: 161039457
Contact Name: James Nichols
Contact Phone: 918 759-7352
Contact
Case #00529239 – Okmulgee county EMS
Status: Closed | Priority: Standard
Q-00030269
**Darrell Hendrix has the parts needed for this service.
This customer has three cot that are in need of a repair. Need work orders generated (see pictures in email)
SN 161039457 the cot is not wanting to lift up
Possible parts needed for repair
16 half shell bearings,