FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
11/14/2025 EMartin sent email to Jason.Johnson@intelashealth.com advising FSE onsite Tues 11/25/25 to complete devices
11/10/2025 MHamersley - recd email reply from Jason, confirming all devices ... 11/10/2025 MHamersley - called (210) 385-0357, spoke to Jason, he will look at email and respond soon ... 10/23/2025 MKlose - called 210-385-0357; straight to VM, box is full; sent email Jason.Johnson@intelashealth.com
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Inspection complete.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed
📞Call Description
FMI 32097
📅
WO-01310585 - FCO - Giraffe OmniBed - SAN ANTONIO - CHILDRENS HOSPITAL OF SAN ANTONIO
Nov 25, 2025 8:00am – 5:00pm
Work Order: WO-01310585
Account: GE Healthcare FMI
Customer Case Number: 1-556244784530
Location: CHILDRENS HOSPITAL OF SAN ANTONIO
Address: 333 N SANTA ROSA ST, SAN ANTONIO, TX 78207
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed
Serial #: TAB24500099SA
Contact Name: Jason Johnson, Director,
PDF
TAB24500099SA_Doors
Nov 25, 2025
PDF
TAB24500099SA_Screws
Nov 25, 2025
PDF
TAB24500099SA_Product Label Pic
Nov 25, 2025
Timeline
📞
Case Linked Oct 21, 12:19 PM
Case Number:00537140
Subject:GE FMI 32097 WO - Christus Santa Rosa
Status:Closed
📋
Work Order Created Oct 21, 3:07 PM
Work Order:WO-01310585
Type:FCO
Priority:PC
🚨15.0 business days
✅
Status: Ready for Release by Mary HamersleyNov 10, 5:29 PM
From:Entered
To:Ready for Release
Duration in Previous:20d 3h
📅
Status: Ready for Scheduling by Mary HamersleyNov 10, 5:40 PM