11/11/25 NFloyd - The original WO-0127744 should have reflected PM on 11 units completed. It was billed for PM on only 7 units at $945 total.
PM has been completed on ALL 14 units originally confirmed plus 1 additional. Billing here for PM on all 8 remaining units to be paid.
10-29 Pgreen per Guldmann customer would like the remaining 3 units PMed and an additional unit. Billing on this WO should only be for 1 unit
🔧Work Performed
Traveled to Hillsboro from Los Angeles
- Performed lift inspection on room 201 - Missing plastic end caps
- Performed lift inspection on room 205 - Found exposed wiring from connection to hand control - Found Broken Plastic on Lift Strap and Plastic Strap Stop on motor - Missing plastic end caps
- Performed lift inspection on room 209 - Found Pin to be missing on end stop(replaced) - Found all wiring to be frayed from battery inside the motor - Found Hanger bar plastic to be damaged - Missing plastic end caps
- Performed lift inspection on room 406 - Found system to be missing all pins on end stops (Only able to install 3 available in stock) - Missing plastic end caps
WO-01311882 - Installation - Hillsboro - Hillsboro Medical Center
Nov 3, 2025 11:00am – 7:00pm
Work Order: WO-01311882
Account: GULDMANN INC
Customer Case Number:
Location: Hillsboro Medical Center
Address: 335 SE 8th Ave, Hillsboro, OR 97123
Manufacturer:
Model #:
Serial #:
Contact Name: Kelly Kintzler
Contact Phone: (800) 664-8834
Contact Email: kki@guldmann.net
Due Date: 11/26/2025
Sch
PDF
Guldmann TBR Hillsboro MC EMSAR
Oct 29, 2025
ZIP
Hillsboro Medial Center
Nov 4, 2025
PDF
Hillsboro Medical Center Service Order
Nov 5, 2025
Timeline
📞
Case Linked Jun 3, 2:37 PM
Case Number:00498498
Subject:Hillsboro Medical Center - RFQ for QTY 14 x GH3+ Inspections -Tech Bid Request