Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Primary Delay34.9 days stuck in "Ready for Release"
⏱️
Total Delay34.7 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
30.8d / 0.5d SLA
Ready for Release
34.9d
Ready for Scheduling
0.1d / 1.0d SLA
Scheduled
5.6d / 2.0d SLA
Tech On Site
1.7d / 1.0d SLA
Completed
0.3d / 1.0d SLA
Closed
1.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/04/2026 MHamersley - recd email from Trai confirming device and ready to schedule for next week ... 02/04/2026 MHamersley - called 913-541-5494, Bio Med Shop, spoke to Trai Gallagher, POC; FSE Cred: VPRO, can use shop if no VPRO; sent f/u email: trai.gallagher@gehealthcare.com - offering date of next week, Feb 9-13 for FSE onsite visit ... 01/13/2026 MHamersley - sent email to William Kasyjanski, Biomedical Technician I; william.kasyjanski@gehealthcare.com
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors
and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
N/A
⚠️Problem Description
Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📅
WO-01311896 - FCO - Giraffe OmniBed Carestation - OVERLAND PARK - OVERLAND PARK REGIONAL
Feb 11, 2026 12:00pm – 5:00pm
Work Order: WO-01311896
Account: GE Healthcare FMI
Customer Case Number: 1-556338223400
Location: OVERLAND PARK REGIONAL
Address: 10500 QUIVIRA RD, OVERLAND PARK, KS 66215
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TABY71375
Contact Name: Trai Gallagher
Conta
PDF
FMI 32097 - Completion Certification Report
Feb 12, 2026
JPG
TABY71375_SN
Feb 12, 2026
Timeline
📞
Case Linked Oct 29, 11:31 AM
Case Number:00539232
Subject:GE FMI 32097 WO - Overland Park
Status:Closed
📋
Work Order Created Oct 29, 1:31 PM
Work Order:WO-01311896
Type:FCO
Priority:PC
🚨31.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:11 PM
From:Entered
To:Ready for Release
Duration in Previous:46d 23h
🚨35.0 business days
📅
Status: Ready for Scheduling by Mary HamersleyFeb 4, 1:06 PM