11/07/2025 EMartin dispatching to FSE onsite per request for SN# H29569721CE01, H302351316C37, H29569705BB95
located at MedOne in Raleigh, NC
🔧Work Performed
H29569705BB95 - System PCBA replacement needed. Pilot sensor zero (0) reading during testing. Multiple sensor errors. Software 1.05.10.00 reloaded. System placed in ship mode. Three-way and proportional valves reseated and torqued. Device reassembled and performed a full diagnostic test. DUT continues to fail pilot [System PCBA sensor] reading. Incident case open, and system PCBA has been added to quote for approval.
Hi, @Erin Martin. This is the work order requested per the below. Thanks!
WO-13895776
Local POC Name (Secondary): Erin Martin
Med One Group - Raleigh
3811 Tarheel Drive Suite 109
Raleigh, NC 27609
📅
WO-01314840 - FCO - Trilogy EV300 Vent - Raleigh - Med One Group - Raleigh
Nov 12, 2025 8:00am – 5:00pm
Work Order: WO-01314840
Account: PHILIPS HEALTHCARE
Customer Case Number: WO-13895776
Location: Med One Group - Raleigh
Address: 3811 Tarheel Drive Suite 109, Raleigh, NC 27609
Manufacturer: Philips Healthcare
Model #: Trilogy EV300 Vent
Serial #: H29569721CE01
Contact Name: Chad Agliam
Contact Phon