Follow up to WO-01295778 * no quote needed per SMC
Closed
Follow up to WO-01295778
Inspection #: 29152
Asset ID: 12939352
Model: Welch Allyn - GS Exam Light IV
Notes: Completed annual inspection, system passed. All Operational and Functional Tests Passed and Returned to Service. Base needs to be attached to light. Out of tolerance.
The technician will need to evaluate light for repair and possible parts needed.
Inspection #: 29152
Asset ID: 12939369
Model: Welch Allyn - GS Exam Light IV
Notes: Completed annual inspection, system passed. All Operational and Functional Tests Passed and Returned to Service. Inspect chassis and A.C. cord for damage. Need to attached base to light.
The technician will need to evaluate light for repair and possible parts needed.
02/11/2026 Jsackett - Deleted SM-248088, returned WO to Billing Correction for Update + Rebill direct to Stony Brook.
12/1 KWeiss, Chatter sent to tech to close w/o for follow up repair case to be created. Can not quote repair from an evaluation w/o.
11/21 KSmith -- Scheduled for 12/1 with Lauren on WO-01316525
11/21 KSmith -- Sent email to schedule for 12/1 on WO-01316525. All FSEs are booked until then. Shawn is aware SLA will be missed
🔧Work Performed
Asset ID: 12939369 - Found short down pipe missing hardware and broken bolt inside threaded pipe. Spoke with Debbie at welch Allyn tech support. Debbie advised there are no sub parts for light stand. Light stand will need to be replaced. Need part #48950 green series exam 4.
🔍Technician Findings
Found light stand separated from base.
⚠️Problem Description
Follow up to WO-01295778
Inspection #: 29152
Inspection #: 29152
Asset ID: 12939369
Model: Welch Allyn - GS Exam Light IV
Notes: Completed annual inspection, system passed. All Operational and Functional Tests Passed and Returned to Service. Inspect chassis and A.C. cord for damage. Need to attached base to light.
The technician will need to evaluate light for repair and possible parts needed.
https://www.ebiotrack.com/workorder.php?wo=69857
Dec 1, 2025 9:21pm
@Mohamed Shareed You will need to close this w/o for a follow up repair case to be created. We can not quote a repair from an evaluation w/o.