FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
Primary Delay36.9 days stuck in "Ready for Release"
⏱️
Total Delay13.5 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
9.0d / 0.5d SLA
Ready for Release
36.9d
Ready for Scheduling
6.0d / 1.0d SLA
Scheduled
0.0d / 2.0d SLA
Tech On Site
0.0d / 1.0d SLA
Completed
0.0d / 1.0d SLA
Closed
1.0d / 1.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
02/06/2026 MHamersley - per email w/Fady, this site is confirmed; FSE will begin the week of Feb 16th; University of Kansas Hospital - Bell Tower Hospital, 4000 Cambridge Street, Kansas City, KS
Jeff Freeman POC Phone #: 913.820.4302 ... 02/04/2026 MHamersley - sent f/u email: jfreeman3@kumc.edu ... 01/19/2026 MHamersley - called 913-588-5000, call went to automated Operator who stated offices are closed today for MLK holiday; sent email: jfreeman3@kumc.edu , inquiring if he is POC for several sites for Univ of Kansas; 913.820.4302 Jeff Freeman.
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors
and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
N/A
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📧
Email: Field Action Notice - FMI 32097 - GE Healthcare - Giraffe Omnibed Heater Door - University of Kansas Hospital - Bell Tower Hospital - Kansas City, KS - WO-01318431
Completed Jan 6, 2026
To: jfreeman3@kumc.edu
CC: deon.malloy@emsar.com; central.scheduling@emsar.com; ricky.bailey@emsar.com; cleopold@kumc.edu
BCC: ricky.bailey@emsar.com
Attachment: FMI32097 MIC UMDC English Customer Letter.pdf
Subject: Field Action Notice - FMI 32097 - GE Healthcare - Giraffe Omnibed Heater Door - University of Kansas Hospital - Bell Tower Hospital - Kansas City, KS - WO-01318431
Body:
Hi Jeff,
Thank you for calling back today. As discussed, I am assisting Deon Malloy and we were contacting
📅
WO-01318431 - FCO - Giraffe OmniBed Carestation - Kansas City - University of Kansas Hospital - Bell Tower Hospital
Feb 15, 2026 7:00pm – 7:00pm
Work Order: WO-01318431
Account: GE Healthcare FMI
Customer Case Number: 1-556634961783
Location: University of Kansas Hospital - Bell Tower Hospital
Address: 4000 Cambridge Street, Kansas City, KS 66160
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB24140008SA
Contact
PDF
FMI 32097 Completion Certification Report
Feb 16, 2026
JPG
TAB24140008SA_SN
Feb 16, 2026
Timeline
📞
Case Linked Dec 2, 9:29 AM
Case Number:00547936
Subject:GE FMI 32097 - Kansas
Status:Closed
⏳2.0 business days
📋
Work Order Created Dec 3, 10:22 AM
Work Order:WO-01318431
Type:FCO
Priority:PC
⚠️9.0 business days
✅
Status: Ready for Release by Mandy NewmanDec 15, 12:10 PM
From:Entered
To:Ready for Release
Duration in Previous:12d 1h
🚨37.0 business days
📅
Status: Ready for Scheduling by Mary HamersleyFeb 6, 10:57 AM