2x Evaluation WOs request for the below units. Repair WOs already created with part orders on them, but customer would like a tech to come onsite to evaluate units while waiting on parts.
Room:Carla Espinosa
Serial Number: v2213129
Midmark 626
Asset ID: 12977373
Table is not syncing to the remotes. They have tried unplugging the beds and syncing them but they are not working.
Room:House
Serial Number: v2213165
Midmark 626
Table is not syncing to the remotes. They have tried unplugging the beds and syncing them but they are not working.
https://www.ebiotrack.com/workorder.php?wo=70392
12-16 Mburris called POC and set appt 12-19 10am ( Edgar Chavez )
🔧Work Performed
Tested the handheld remote and confirmed it is defective and requires replacement. Contacted Midmark for the part number and was advised the correct replacement is P/N 002-10049-02. Point of contact (Aseel) indicated they had spare remotes available and provided one. The replacement remote was successfully synced to the exam chair. All functions were tested and are operating normally.
🔍Technician Findings
Arrived on site and confirmed handheld remote for exam chair 626-003 with SN: V2213165 in room "House" is not operating properly. It's only responsive to raising the base function.
⚠️Problem Description
1 O 2
2x Evaluation WOs request for the below units. Repair WOs already created with part orders on them, but customer would like a tech to come onsite to evaluate units while waiting on parts.
Room:Carla Espinosa
Serial Number: v2213129
Midmark 626
Asset ID: 12977373
Table is not syncing to the remotes. They have tried unplugging the beds and syncing them but they are not working.
Room:House
Serial Number: v2213165
Midmark 626
Table is not syncing to the remotes. They have tried unplugging the beds and syncing them but they are not working.
https://www.ebiotrack.com/workorder.php?wo=70392
📧
Email: 193 Jefferson Drive Menlo Park, CA 94025 / Evaluation
Completed Dec 18, 2025
To: edgar.chavez@crossoverhealth.com
CC:
BCC: michael.burris@emsar.com
Attachment: --none--
Subject: 193 Jefferson Drive Menlo Park, CA 94025 / Evaluation
Body:
Hi,
Changed tech to Angel Rios
Best Regards,
Michael Burris Jr
Account Service Coordinator I
Work Phone : (937) 802-9367
michael.burris@emsar.com
www.emsar.com
📅
WO-01326688 - Evaluation - Menlo Park - Crossover Health - Life @ Wellness Center - MPK281
Dec 19, 2025 1:00pm – 3:00pm
Work Order: WO-01326688
Account: Crossover Health Management Services
Customer Case Number:
Location: Crossover Health - Life @ Wellness Center - MPK281
Address: 193 Jefferson Drive, Menlo Park, CA 94025
Manufacturer:
Model #:
Serial #:
Contact Name: Edgar Chavez
Contact Phone: 9252346755
Contac
Case #00550605 – L2 review - HCP
Status: Closed | Priority: High
1 O 2
2x Evaluation WOs request for the below units. Repair WOs already created with part orders on them, but customer would like a tech to come onsite to evaluate units while waiting on parts.
Room:Carla Espinosa
Serial Number: v2213129
Midmark 626
Asset ID: 12977373
Table is not syn