1 O 2
2x Evaluation WOs request for the below units. Repair WOs already created with part orders on them, but customer would like a tech to come onsite to evaluate units while waiting on parts.
Room:Carla Espinosa
Serial Number: v2213129
Midmark 626
Asset ID: 12977373
Table is not syncing to the remotes. They have tried unplugging the beds and syncing them but they are not working.
Room:House
Serial Number: v2213165
Midmark 626
Table is not syncing to the remotes. They have tried unplugging the beds and syncing them but they are not working.
https://www.ebiotrack.com/workorder.php?wo=70392
12-16 Mburris called POC and set appt 12-19 10am ( Edgar Chavez )
🔧Work Performed
The customer provided spare batteries, after which the handheld remote was successfully synced to the exam chair. All functions on the handheld remote are operating normally. The customer inquired about replacing the wireless controller with a wired option. Contacted Midmark technical support, who advised this is possible using the Wired Control Upgrade Kit (P/N 002-10369-00). The customer is satisfied with the wireless control at this time but may consider upgrading to a wired control if issues persist.
🔍Technician Findings
Arrived on site and confirmed handheld remote for exam chair 626-003 with SN: V2213129 in room "Carla Espinosa" is not operating properly. It's completely unresponsive.
⚠️Problem Description
2 O 2
2x Evaluation WOs request for the below units. Repair WOs already created with part orders on them, but customer would like a tech to come onsite to evaluate units while waiting on parts.
Room:Carla Espinosa
Serial Number: v2213129
Midmark 626
Asset ID: 12977373
Table is not syncing to the remotes. They have tried unplugging the beds and syncing them but they are not working.
Room:House
Serial Number: v2213165
Midmark 626
Table is not syncing to the remotes. They have tried unplugging the beds and syncing them but they are not working.
https://www.ebiotrack.com/workorder.php?wo=70393
📋
Outbound to +19252346755
Completed Dec 16, 2025
12-16 Mburris called POC and set appt 12-19 10am ( Edgar Chavez )
📅
WO-01326689 - Evaluation - Menlo Park - Crossover Health - Life @ Wellness Center - MPK281
Dec 19, 2025 1:00pm – 3:00pm
Work Order: WO-01326689
Account: Crossover Health Management Services
Customer Case Number:
Location: Crossover Health - Life @ Wellness Center - MPK281
Address: 193 Jefferson Drive, Menlo Park, CA 94025
Manufacturer:
Model #:
Serial #:
Contact Name: Edgar Chavez
Contact Phone: 9252346755
Contac
Case #00550605 – L2 review - HCP
Status: Closed | Priority: High
1 O 2
2x Evaluation WOs request for the below units. Repair WOs already created with part orders on them, but customer would like a tech to come onsite to evaluate units while waiting on parts.
Room:Carla Espinosa
Serial Number: v2213129
Midmark 626
Asset ID: 12977373
Table is not syn