FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
01/07/2026 DMalloy - Email recieved from customer Confirming the devices. Replied advising that I will have FSE Jackson in the area on next week Thursday 1/15/2026. calvin.crum@uky.edu
01/07/2026 MHamersley - recd email from Clavin Crum, POC, 859-297-8332, confirming devices at UK ALBERT B CHANDLER HOSPITAL, ATTN: Clinical Engineering, Room H0080, 800 Rose Street Lexington, KY 40536; FSE Cred: Vendormate; ... 01/07/2026 MHamersley - called 859-218-5114, call went to VM/LM; sent email: calvin.crum@uky.edu
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors
and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Device serviced.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
📄Description
GE Healthcare Giraffe OmniBed
📞Call Description
FMI 32097
📅
WO-01332989 - FMI 32097 - Giraffe OmniBed - Lexington - UK ALBERT B CHANDLER HOSPITAL
Jan 15, 2026 8:00am – 5:00pm
Work Order: WO-01332989
Account: GE Healthcare FMI
Customer Case Number: 1-556758886525
Location: UK ALBERT B CHANDLER HOSPITAL
Address: 800 Rose Street, Lexington, KY 40536
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed
Serial #: TAB24240049SA
Contact Name: Calvin Crum
Contact Phone:
JPEG
TAB24240049SA
Jan 15, 2026
PDF
Checklist Lexington#1
Jan 15, 2026
Timeline
📞
Case Linked Jan 6, 1:43 PM
Case Number:00556459
Subject:GE FMI 32097 - UK Healthcare
Status:Closed
📋
Work Order Created Jan 6, 4:27 PM
Work Order:WO-01332989
Type:FMI 32097
Priority:PC
📅
Status: Ready for Scheduling by Mary HamersleyJan 7, 4:25 PM