Primary Delay16.0 days stuck in "Ready for Scheduling"
⏱️
Total Delay19.8 business days over SLA thresholds
SLA vs Actual Time by Status
Actual (within SLA)
Actual (over SLA)
SLA Target
Entered
0.0d / 0.5d SLA
Ready for Scheduling
16.0d / 1.0d SLA
Scheduled
5.8d / 2.0d SLA
Completed
0.0d / 1.0d SLA
Closed
2.0d / 1.0d SLA
Ready to Bill
0.8d / 1.0d SLA
Invoice Pending
0.0d / 2.0d SLA
💬 Activities, Notes & Communications
🔒Internal CSA Notes
06/23/2025 RArnold - Dispatching to Dillon.
06/02/2025 RArnold - Dispatching to Dillon for today.
🔧Work Performed
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
🔍Technician Findings
Template Name: FMI32097_Code 1_inspect.
Template Description: FMI32097_Code 1_inspect.
Problem Found: FMI32097_Heater door inspection.
Action Taken: FMI 32097 completed. Inspected the Heater doors and confirmed screws tightened to the specification.
Verification Test:
1. Verified the canopy moves smoothly.
2. Verified transition mode message is not present.
3. Verified doors close smoothly.
4. Verified corner pins are aligned.
5. Verified no alarms.
The device has passed all tests and is ready for customer use.
⚠️Problem Description
FMI 32097:- Inspection of heater door screws for Giraffe OmniBed and Giraffe OmniBed Carestation Purpose:- Certain Giraffe OmniBed and Giraffe OmniBed Carestation may have defective heater doors due to improper torquing of heater door screws. FMI Type: Mandatory Recall FMI For code 2, 3, 4 and 5 closures, please reach out to FMI admin team at tcs.fmiadmins@gehealthcare.com or 8668836784 – OPT. 5 for assistance in closing and correct objective evidence attachment to Activity/SR.
SN: TAB24190056SA
📄Description
GE Healthcare Giraffe OmniBed Carestation CS1
📞Call Description
FMI 32097
📅
WO-01276389 - FCO - Giraffe OmniBed Carestation - NEW YORK - NYC HEALTH + HOSPITALS BELLEVUE
Jun 23, 2025 8:00am – 5:00pm
Work Order: WO-01276389
Account: GE Healthcare FMI
Customer Case Number: 1-554958495143
Location: NYC HEALTH + HOSPITALS BELLEVUE
Address: 462 1ST AVE, NEW YORK, NY 10016
Manufacturer: GE Healthcare
Model #: Giraffe OmniBed Carestation
Serial #: TAB24190056SA
Contact Name: Fakhrun Nabi
Contact Phone
JPG
TAB24190056SA DOOR OPEN
Jun 30, 2025
JPG
TAB24190056SA RIGHT SIDE SCREW
Jun 30, 2025
JPG
TAB24190056SA LEFT SIDE SCREW
Jun 30, 2025
JPG
TAB24190056SA DOOR CLOSED
Jun 30, 2025
JPG
GIRAFFE OMNIBED CARESTATION TAB24190056SA
Jun 30, 2025
PDF
FMI Completion Certification Report TAB24190056SA
Jun 30, 2025
Timeline
📞
Case Linked May 29, 4:47 PM
Case Number:00497326
Subject:Urgent GE FMI WO request. Waiting for address to location